SIGLENT
94 SSA3000X Plus Programming Guide
Return
Default
1
Menu
Trace > Num of Traces
Example
:CALCulate:PARameter:COUNt 4
Command
Format
:DISPlay:LAYout <integer,integer>
Instruction
Sets trace layout on screen
Currently, one row, two columns are not supported (1, 2)
Parameter
Type
Integer (rows, columns)
Parameter
Range
rows 1 ~ 2
columns 1 ~ 2
Return
Default
two rows, two columns
Menu
Trace > Layout
Example
:DISPlay:LAYout 2,2
Command
Format
:TRACe[1]|2|3|4:DATA:NAME
:TRACe[1]|2|3|4:DATA:NAME?
Instruction
Sets trace format.
Gets trace format.
Parameter
Type
Enumeration
Parameter
Range
TIME: time
SPECtrum: spectrum
MTIMe: IQ meas time
MSPEctrum: IQ meas spectrum (FFT of IQ Meas Time.)
RTIMe: IQ Reference time (Reconstructed ideal time waveform to compare
IQ Meas Time against)
RSPEctrum: IQ Reference spectrum(FFT of IQ Reference time.)
MERRor: IQ Mag Err (Difference in length of the IQ Meas Time vector and IQ
Ref Time vectorat each point in time.)
PERRor: IQ Phase Err (Difference in phase of the IQ Meas Time vector and
IQ Ref Time vector at each point in time.)
EVTime: Error Time (Vector difference between IQ Meas Time and IQ Ref
Time at each pointin time.)
EVSPectrum: Error Vector Spec