Programming Guide
Table Of Contents
- 1. Programming Overview
- 2. SCPI Overview
- 3. Commands that are Common to All Modes
- 4. Spectrum Analyzer
- 5. Modulation Analyzer
- 6. Programming Examples
SIGLENT
92 SSA3000X Plus Programming Guide
Parameter
Type
Enumeration
RECTangular
HAMMing :
HANNing
FLATtop
BLACkman
Parameter
Range
None
Return
Enumeration RECT
HAMM
HANN
FLAT
BLAC
Default
100 kHz
Menu
BW > Window
Example
:DDEMod:FFT:WINDow:TYPE BLAC
5.4 Sweep Subsection
:INITiat[:IMMediate]
:INITiate:CONTinuous
ABORt
Command
Format
:INITiat[:IMMediate]
Instruction
Restart the current sweep.
:INITiate:RESTart and :INITiate:IMMediate perform exactly the same function.
Parameter
Type
None
Parameter
Range
None
Return
None
Default
None
Menu
Example
:INITiate:IMMediate
Command
Format
:INITiate:CONTinuous OFF|ON|0|1
:INITiate:CONTinuous?
Instruction
Sets continuous sweep mode on-off.
Gets continuous sweep mode state.
Parameter
Type
Boolean
Parameter
Range
OFF|ON|0|1
Return
0|1