SIGLENT
88 SSA3000X Plus Programming Guide
5.Modulation Analyzer
7.1 Frequency Subsection ............................................. 错误!未定义书签。
7.2 Amplitude Subsection ............................................. 错误!未定义书签。
7.3 BW Subsection ......................................................... 错误!未定义书签。
7.4 Sweep Subsection.................................................... 错误!未定义书签。
7.5 Trace Subsection...................................................... 错误!未定义书签。
7.6 Marker Subsection ................................................... 错误!未定义书签。
7.7 Measurement Subsystem ........................................ 错误!未定义书签。
7.8 Trigger Subsection ................................................... 错误!未定义书签。
5.1 Frequency Subsection
[:SENSe]:FREQuency:CENTer
[:SENSe]:FREQuency:CENTer:STEP[:INCRement]
[:SENSe]:FREQuency:SPAN?
Command
Format
[:SENSe]:FREQuency:CENTer <freq>
[:SENSe]:FREQuency:CENTer?
Instruction
Sets the center frequency
Gets the center frequency.
Parameter
Type
Float, unit: Hz, kHz, MHz, GHz
Parameter
Range
0 Hz ~ 3.2 GHz
Return
Float, unit: Hz
Default
100 MHz
Menu
Frequency > Center Freq
Example
[:SENSe]:FREQuency:CENTer 300 MHz
Command
Format
[:SENSe]:FREQuency:CENTer:STEP[:INCRement] <freq>
[:SENSe]:FREQuency:CENTer:STEP[:INCRement]?
Instruction
Sets frequency step
Gets frequency step
Parameter
Type
Float, unit: Hz, kHz, MHz, GHz
Parameter