SSA3000X Plus Programming Guide 65
Range
Return
0|1
Default
OFF
Menu
Limit > Limit1|2
Example
:CALCulate:LLINe1:STATe OFF
Command
Format
:CALCulate:LLINe[1]|2:TYPE UPPer|LOWer
:CALCulate:LLINe[1]|2:TYPE?
Instruction
Mode sets a limit line to be either an upper or lower type limit line. An upper
line will be used as the maximum allowable value when comparing with the
data.
Gets limit type.
Parameter
Type
Enumeration
Parameter
Range
UPPer|LOWer
Return
Enumeration
Default
The default setting of LINe1 is UPPer, the default setting of LINe2 is LOWer
Menu
Limit > Limit1|2 Edit > Type
Example
:CALCulate:LLINe1: TYPE LOWer
Command
Format
:CALCulate:LLINe[1]|2:MODE LINE|POINt
:CALCulate:LLINe[1]|2:MODE?
Instruction
Sets limit mode
Gets limit mode
Parameter
Type
Enumeration
Parameter
Range
LINE|POINt
Return
Enumeration
Default
LINE
Menu
Limit > Limit1|2 Edit > Mode
Example
:CALCulate:LLINe1: MODE POINt
Command
Format
:CALCulate:LLINe[1]|2:Y <value>
:CALCulate:LLINe[1]|2:Y?
Instruction
Sets the Y-axis value of a limit line. Limit line Y-axis value is set
independently and is not affected by the X-axis units.
Gets the Y-axis value of a limit line.
Parameter
Type
Float
Parameter
-400 dBm~330 dBm