Programming Guide
Table Of Contents
- 1. Programming Overview
- 2. SCPI Overview
- 3. Commands that are Common to All Modes
- 4. Spectrum Analyzer
- 5. Modulation Analyzer
- 6. Programming Examples
SIGLENT
SSA3000X Plus Programming Guide 63
Default
FREQuency
Menu
Marker Fn > Read Out
Example
:CALCulate:MARKer1:X:READout FREQuency
Command
Format
:CALCulate:MARKer:TRCKing[:STATe] OFF|ON|0|1
:CALCulate:MARKer:TRCKing[:STATe]?
Instruction
This command turns on/off signal track state.
Gets signal track state.
Parameter
Type
Boolean
Parameter
Range
OFF|ON|0|1
Return
0|1
Default
ON
Menu
Frequency > Signal Track
Example
:CALCulate:MARKer:TRCKing on
4.8 Limit Subsection
:CALCulate:LLINe:TEST:STARt
:CALCulate:LLINe:TEST:STOP
:CALCulate:LLINe:TEST:STATe?
:CALCulate:LLINe[1]|2:STATe
:CALCulate:LLINe[1]|2:TYPE
:CALCulate:LLINe[1]|2:MODE
:CALCulate:LLINe[1]|2:Y
:CALCulate:LLINe[1]|2:DATA
:CALCulate:LLINe[1]|2:ADD
:CALCulate:LLINe[1]|2:DELete
:CALCulate:LLINe[1]|2:ALL:DELete
:CALCulate:LLINe:CONTrol:DOMain
:CALCulate:LLINe:CONTrol:BEEP
:CALCulate:LLINe:FAIL?
:CALCulate:LLINe:FAIL:STOP
Command
Format
:CALCulate:LLINe:TEST:STARt