Programming Guide
Table Of Contents
- 1. Programming Overview
- 2. SCPI Overview
- 3. Commands that are Common to All Modes
- 4. Spectrum Analyzer
- 5. Modulation Analyzer
- 6. Programming Examples
SIGLENT
48 SSA3000X Plus Programming Guide
Example
:TRACe:MATH:X A
Command
Format
:TRACe:MATH:Y A|B|C|D
:TRACe:MATH:Y?
Instruction
Sets trace math input Y
Gets trace math input Y
Parameter
Type
Enumeration
Parameter
Range
A|B|C|D 或 TRACE1|TRACE2|TRACE3|TRACE4
Return
Enumeration
Default
A
Menu
Trace > Math > Input Y
Example
:TRACe:MATH:Y A
Command
Format
:TRACe:MATH:Z A|B|C|D
:TRACe:MATH:Z?
Instruction
Sets trace math Output Z
Gets trace math Output Z
Parameter
Type
Enumeration
Parameter
Range
A|B|C|D 或 TRACE1|TRACE2|TRACE3|TRACE4
Return
Enumeration
Default
A
Menu
Trace > Math > Output Z
Example
:TRACe:MATH:Z A
Command
Format
:TRACe:MATH:OFFSet <const>
:TRACe:MATH:OFFSet?
Instruction
Sets trace math OFFSet
Gets trace math OFFSet
Parameter
Type
Float
Parameter
Range
-100 dB ~100 dB
Return
Float
Default
0.00 dB
Menu
Trace > Math > Offset
Example
:TRACe:MATH:OFFSet 7