SSA3000X Plus Programming Guide 33
[:SENSe]:POWer[:RF]:ATTenuation:AUTO
[:SENSe]:POWer[:RF]:GAIN[:STATe]
:DISPlay:WINDow:TRACe:Y:SCALe:RLEVel:OFFSet
:UNIT:POWer
:DISPlay:WINDow:TRACe:Y[:SCALe]:SPACing
:DISPlay:WINDow:TRACe:Y[:SCALe]:PDIVision
[:SENSe]:CORRection:OFF
[:SENSe]:CORRection:CSET:ALL[:STATe]
[:SENSe]:CORRection:CSET[1]|2|3|4[:STATe]
[:SENSe]:CORRection:CSET[1]|2|3|4:ADD
[:SENSe]:CORRection:CSET[1]|2|3|4:DELete
[:SENSe]:CORRection:CSET[1]|2|3|4:ALL:DELete
[:SENSe]:CORRection:CSET[1]|2|3|4:DATA
[:SENSe]:CORRection:IMPedance[:INPut][:MAGNitude]
Command
Format
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel <value>
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel?
Instruction
This command sets the reference level for the Y-axis.
Gets reference level.
Parameter
Type
Float, unit: dBm, dBmV, dBuV, V, W
Parameter
Range
Unit is dBm: -100 dBm ~ 30 dBm
Unit is dBmV: -53.01 dBmV ~ 76.99 dBmV,
Unit is dBuV: 6.99 dBuV ~ 136.99 dBuV,
Unit is Volts: 2.24 uV ~ 7.07 V
Unit is Watts: 100 fW ~ 1 W
Return
Float, unit: dBm
Default
0 dBm
Menu
Amplitude > Ref Level
Example
:DISPlay:WINDow:TRACe:Y:RLEVel 20 DBM
Command
Format
[:SENSe]:POWer[:RF]:ATTenuation <value>
[:SENSe]:POWer[:RF]:ATTenuation?
Instruction
Sets the input attenuator of the spectrum analyzer.
Gets the input attenuator.
Parameter
Type
Integer
Parameter
Range
0 dB ~ 51 dB
Return
Integer, unit: dB