Programming Guide
Table Of Contents
- 1. Programming Overview
- 2. SCPI Overview
- 3. Commands that are Common to All Modes
- 4. Spectrum Analyzer
- 5. Modulation Analyzer
- 6. Programming Examples
SIGLENT
18 SSA3000X Plus Programming Guide
Example
*OPC?
Command
Format
*SRE <integer>
*SRE?
Instruction
This command enables the desired bits of the service request enable
register.
The query returns the value of the register, indicating which bits are currently
enabled. The default value is 255.
Menu
None
Example
*SRE 1
Command
Format
*STB
Instruction
This query is used by some instruments for a self test.
Menu
None
Example
*STB
Command
Format
*WAI
Instruction
This command causes the instrument to wait until all pending commands are
completed before executing any additional commands.
There is no query form to the command.
Menu
None
Example
*WAI
Command
Format
*TRG
Instruction
Restart the current sweep.
Menu
None
Example
*TRG
Command
Format
*TST?
Instruction
This query is used by some instruments for a self test.
Menu
None
Example
*TST?