SSA3000X Plus Programming Guide 103
Parameter
Range
1 ~ 1000
Return
Integer
Default
10
Menu
Meas> Statistic > Avg
Example
:AVERage:COUNt 20
Command
Format
:CALCulate:RESTart
Instruction
Restart measurements
Parameter
Type
None
Parameter
Range
None
Return
None
Default
None
Menu
Meas > Statistic > Restart Meas
Example
:CALCulate:RESTart
Command
Format
:READ:DDEMod?
Instruction
Read digital demod result
If demod type is ASK it will return:
1. ASK err rms (% rms)
2. ASK err peak (% pk)
3. symbol position of ASK err peak
4. carrier power
5. carrier offset
6. ASK depth
If demod type is FSK it will return:
1.FSK err rms (% rms)
2.FSK err peak (% pk)
3.symbol position of FSK err peak
4. carrier power
5.carrier offset
6.FSK deviation
If If demod type is MSK,PSK,QAM it will return:
1. EVM rms (% rms)
2. EVM peak (% pk)
3. symbol position of EVM peak
4.magnitude error rms (% rms).
5. magnitude error peak (% pk)
6.symbol position of magnitude error peak
7.phase error rms (deg)
8.phase error peak (deg pk)
9.symbol position of phase error peak