SSA3000X Plus Programming Guide 93
ON
Sweep > Sweep
:INITiate:CONTinuous OFF
ABORt
This command is used to stop the current measurement. It aborts the
currentmeasurement as quickly as possible, resets the sweep and trigger
systems, and puts the measurement into an "idle" state.
If the analyzer is set for Continuous measurement, it sets up the
measurement and initiates a new data measurement sequence with a new
data acquisition (sweep)taken once the trigger condition is met.
If the analyzer is set for Single measurement, it remains in the "idle" state
until an :INIT:IMM command is received.
None
None
None
None
None
INIT;ABORt
5.5 Trace Subsection
:CALCulate:PARameter:COUNt
:DISPlay:LAYout
:TRACe[1]|2|3|4:DATA:NAME
:TRACe[1]|2|3|4:FORMat[:Y]
:TRACe:COPY
:TRACe:DEMod:EYE:LENGth
:TRACe:DEMod:TABLe:FORMat
:CALCulate:PARameter:COUNt <integer>
:CALCulate:PARameter:COUNt?
Sets trace number.
Integer
1 ~ 4