Programming Guide
Table Of Contents
- 1. Programming Overview
- 2. SCPI Overview
- 3. Commands that are Common to All Modes
- 4. Spectrum Analyzer
- 5. Modulation Analyzer
- 6. Programming Examples
SIGLENT
SSA3000X Plus Programming Guide 89
Range
Return
Float, unit: Hz
Default
10 kHz
Menu
Frequency > Freq Step
Example
[:SENSe]:FREQuency:CENTer:STEP[:INCRement] 20 MHz
Command
Format
[:SENSe]:FREQuency:SPAN?
Instruction
Query span
The span of modulation analyzer mode is determined by multiple
measurement parameters, and can not be set directly.
Parameter
Type
None
Parameter
Range
None
Return
Float, unit: Hz
Default
31.25 kHz
Menu
Span > Span
Example
[:SENSe]:FREQuency:SPAN?
5.2 Amplitude Subsection
[:SENSe]:POWer[:RF]:ATTenuation
[:SENSe]:POWer[:RF]:ATTenuation:AUTO
:TRACe1|2|3|4:Y[:SCALe]:RLEVel
:TRACe1|2|3|4:Y[:SCALe]:PDIVision
:TRACe1|2|3|4[:Y]:AUToscale
Command
Format
[:SENSe]:POWer[:RF]:ATTenuation <value>
[:SENSe]:POWer[:RF]:ATTenuation?
Instruction
Sets the input attenuator
Gets the input attenuator
Parameter
Type
Integer, Unit: dB
Parameter
Range
0 dB ~ 51 dB
Return
Integer, unit: dB
Default
20 dB
Menu
Amplitude > Attenuator
Example
[:SENSe]:POWer[:RF]:ATTenuation 30 dB