Programming Guide
Table Of Contents
- 1. Programming Overview
- 2. SCPI Overview
- 3. Commands that are Common to All Modes
- 4. Spectrum Analyzer
- 5. Modulation Analyzer
- 6. Programming Examples
SIGLENT
82 SSA3000X Plus Programming Guide
Instruction
Set Frequency Step
Get Frequency Step
Parameter
Type
Float, Unit: Hz, kHz, MHz, GHz
Parameter
Range
10M Hz ~ 3.19 GHz
Return
Float, Unit:Hz
Default
Menu
Meas > Harmoniscs > Freq Step
Example
:HARMonics:FREQuency:STEP 1 MHz
Command
Format
[:SENSe]:HARMonics:FREQuency:FUNDamental:AUTO
[:SENSe]:HARMonics:FREQuency:FUNDamental:AUTO?
Instruction
Sets Fundamental Freq State
Gets Fundamental Freq State
Parameter
Type
Boolean
Parameter
Range
OFF|ON|0|1
Return
Boolean
Default
1
Menu
Meas > Harmonics > Fundamental
Example
:HARMonics:FREQuency:FUNDamental:AUTO on
Command
Format
[:SENSe]:HARMonics:FREQuency:STEP[:INCRement]:AUTO
[:SENSe]:HARMonics:FREQuency:STEP[:INCRement]:AUTO?
Instruction
Sets Freq step State
Gets Freq step State
Parameter
Type
Boolean
Parameter
Range
OFF|ON|0|1
Return
Boolean
Default
1
Menu
Meas > Harmonics > Freq Step
Example
:HARMonics:FREQuency:STEP:AUTO on
Command
Format
[:SENSe]:HARMonics:NUMBer
[:SENSe]:HARMonics:NUMBer?
Instruction
Sets Harmonic Num
Gets Harmonic Num
Parameter
Type
Integer