Programming Guide

Table Of Contents
SIGLENT
24 SSA3000X Plus Programming Guide
Instruction
Sets both the measure and setting parameters to factory preset parameters.
Parameter
Type
None
Parameter
Range
None
Return
None
Default
None
Menu
System > Pwr/Preset > Factory Reset
Example
:SYSTem:FDEFault
Command
Format
:SYSTem:LKEY <“option”>,<“license key”>
Instruction
Use this command to enable the specified option with the license key, please
restart the instrument to make license active.
Parameter
Type
“option”: Enumeration
“ license key”: String
Parameter
Range
“option”: Meas|EMI|TG|DMA|
“ license key”: provided by Siglent Technologies, 16 bits String.
Return
None
Default
None
Menu
System > System Info > Load Option
Example
:SYSTem:LKEY EMI,fjbdajffnklmgwno
Command
Format
:SYSTem:OPTions?
Instruction
This command returns a list of the options that are installed.
Parameter
Type
None
Parameter
Range
None
Return
Meas|EMI|TG|DMA
Default
None
Menu
System > System Info
Example
:SYSTem:OPTions?
Command
Format
:SYSTem:POWer:OFF
Instruction
Use this command to turn off the instrument.
Parameter
Type
None
Parameter
None