SAP1000 Active Probe User Manual
9 / 19 WW W.SIGLENT.COM
2. Connect the SAP1000 probe to channel 1 of the oscilloscope.
3. Touch Probe in the C1 menu, and then check the information of the probe,
including the probe type, probe SN, bandwidth, resistance, capacitance
and attenuation.
4. Set the vertical scale of channel 1 to 2 V/div.
5. Set the offset of channel 1 to zero.
6. Measure the average voltage of channel 1.The measurement should be
within ± (1.5% * full screen reading + 10 mV).
7. Set the vertical scale of channel 1 to1 V/div, 500 mV/div, 200 mV/div, 100
mV/div, 50 mV/div, 20 mV/div, 10 mV/div, and repeat step 6.