User Manual

SDS1000X-U Series Digital Oscilloscope
14
WWW.SIGLENT.COM
Language
Simplified Chinese, Traditional Chinese, English, French, Japanese, Korean, German, Russian,
Italian, Portuguese
Environments
Temperature
Operating: 0 - +40
Non-operating: -20 - + 60
Humidity
Operating: 85% RH, 40 , 24 hours
Non-operating: 85% RH, 65 , 24 hours
Height
Operating: 3000 m
Non-operating: 15,000 m
Standards
Electromagnetic
compatibility
Meets EMC directive (2014/30/EU), meets or exceeds IEC 61326-1:2012/EN61326-1:2013 (Basic)
Conducted disturbance
CISPR 11/EN 55011
CLASS A group 1 150kHz-
30MHz
Radiated disturbance
CISPR 11/EN 55011
CLASS A group 1 30MHz-
1GHz
Electrostatic discharge (ESD)
IEC 61000-4-2/EN 61000-4-2
4.0 kVContact), 8.0 kVAir
Radio-frequency
electromagnetic field
Immunity
IEC 61000-4-3/EN 61000-4-3
10 V/m80 MHz to 1 GHz;
3 V/m1.4 GHz to 2 GHz;
1 V/m2.0 GHz to 2.7GHz
Electrical fast transients (EFT)
IEC 61000-4-4/EN 61000-4-4
2kV Input AC Power Ports
Surges
IEC 61000-4-5/EN 61000-4-5
1kV Line to line
2kV Line to ground
Radio-frequency continuous
conducted Immunity
IEC 61000-4-6/EN 61000-4-6
3 V, 0.15-80MHz
Voltage dips and
interruptions
IEC 61000-4-11/EN 61000-4-
11
Voltage Dips
0% UT during 1 cycle;
40% UT during 10/12 cycles;
70% UT during 25/30 cycles
Voltage interruptions0% UT
during 250/300 cycles
Safety
UL 61010-1:2012/R: 2018-11; CAN/CSA-C22.2 No. 61010-1:2012/A1:2018-11.
UL 61010-2-030:2018; CAN/CSA-C22.2 No. 61010-2-030:2018.
Power Supply
Input Voltage
100 240 Vrms 50/60Hz
100 120 Vrms 400Hz
Power
50 W Max