Programming Guide

Digital Oscilloscopes Series
95
FUNCTION
PF_SET,PFST
Command /Query
DESCRIPTION
The PF_SET command sets the X mask and the Y
mask of the mask setting in the pass/fail option.
The response to the PF_ SET? query indicates the
value of the X mask and the Y mask.
COMMAND SYNTAX
PF_ SET XMASK, <div>,YMASK, <div>
<div> : = 0.04div~4.0div
QUERY SYNTAX
PF_ SET?
RESPONSE FORMAT
PF_ SET XMASK, <div>,YMASK, <div>
EXAMPLE
The following instruction sets the X mask to 0.4div
and the Y mask to 0.5div of the mask setting in the
pass/fail option:
Command message:
PFST XMASK,0.4,YMASK,0.5
RELATED COMMANDS
PFSL PFST