Programming Guide

Digital Oscilloscopes Series
38
CURSOR
CURSOR_SET, CRST
Command /Query
DESCRIPTION
The CURSOR_SET command allows the user to
position any one of the eight independent cursors at
a given screen location. The positions of the cursors
can be modified or queried even if the required
cursor is not currently displayed on the screen.
When setting a cursor position, a trace must be
specified, relative to which the cursor will be
positioned.
The CURSOR_SET? Query indicates the current
position of the cursor(s). The values returned
depend on the grid type selected.
Notation
VREF
The volt-value of curA under manual cursor mode
VDIF
The volt -value of curB under manual cursor mode
TREF
The time value of curA under manual cursor mode
TDIF
The time value of curB under manual cursor mode
HREF
The time value of curA under Track cursor mode
HDIF
The time value of curB under Track cursor mode
Notation
VREF
The volt-value of curA under manual cursor mode
VDIF
The volt -value of curB under manual cursor mode
TREF
The time value of curA under manual cursor mode
TDIF
The time value of curB under manual cursor mode
HREF
The time value of curA under Track cursor mode
Notation
VREF
The voltage-value of curA under
manual cursor mode
VDIF
The voltage-value of curB under
manual cursor mode
TREF
The time value of curA under
manual cursor mode
TDIF
The time value of curB under
manual cursor mode
HREF
The time value of curA under Track
cursor mode
HDIF
The time value of curB under Track
cursor mode
COMMAND SYNTAX
<trace>:CuRsor_SeT<cursor>,<position>[,<cursor
>,<position>,<cursor> ,<position>]
< trace > : = {C1, C2, C3, C4}
<cursor> : =
{ VREF,VDIF,TREF,TDIF,HRDF,HDIF}
<position>(horizontal)= {0.02 to 13.98 DIV }
<position>(vertical)= {0.02 to 7.98 DIV}
Note:
1. The horizontal positions value is related to the
size of screen. For SPO models, the positions
value is in the range of 0.02 to 13.98. And in
non-SPO models its in the range of -8 to 8. If the
value is in the range of 0.02 to 13.98, you need