Programming Guide
Digital Oscilloscopes Series
112
WAVEFORM TRANSFER
STORE, STO
Command
DESCRIPTION
The STORE command stores the contents of
the specified trace into the current directory in
a USB memory device.
COMMAND SYNTAX
STOre <trace>
<trace>: = {TA, TB, TC, TD, C1, C2, C3,
C4,ALL_DISPLAYED}
<dest>: = { UDSK}
Note:
1. If the STORE command is sent without any
argument, and the current trace isn‘t
enabled, the current trace will be enabled
and stored in the Store Setup. This setup can
be modified using the STORE_SETUP
command.
2. The <dest> parameter is vary with
oscilloscope‘s models. If your oscilloscope
is non-SPO models , this parameter‘s
options are {M1~M10(or M20 in the CFL
series)}, UDSK}. If your oscilloscope is
SPO model, this parameter‘s option is
{UDSK}.
EXAMPLE
The following command stores the contents of
Channel 1(C1) into USB memory device:
Command message:
STO C1, UDSK
The following command stores all currently
displayed waveforms onto the USB memory
device:
Command message:
STO ALL_DISPLAYED, UDSK
The following command stores the contents of
Channel 1(C1) into Memory 1 (M1):
Command message: