Data Sheet
WWW.SIGLENT.COM
SDS1000DL+/CML+ Series Digital Oscilloscope
Horizontal System
Timebase Scale
150 MHz 2.5 ns/div - 50 s/div
100 MHz 2.5 ns/div - 50 s/div
70 MHz 5.0 ns/div - 50 s/div
50 MHz 5.0 ns/div - 50 s/div
Channel Skew <500 ps
Display Format Y-T, X-Y, Scan
Timebase Accuracy ±50 ppm
Scan Mode 100 ms/div ~ 50 s/div
Vertical System
Bandwidth (-3 dB)
150 MHz (SDS1152 CML+)
100 MHz (SDS1102 CML+)
70 MHz (SDS1072 CML+)
50 MHz (SDS1052 DL+)
Vertical Resolution 8 bit
Vertical Scale (Probe 1 X) 2 mV/div - 10 V/div (1-2-5 )
Offset Range (Probe 1 X) 2 mV - 200 mV: ± 1.6 V; 206 mV ~ 10 V: ± 40 V
Bandwidth Limit 20 MHz ±40%
Bandwidth Flatness
DC - 10%(BW): ± 1 dB
10% - 50%(BW): ± 2 dB
50% - 100%(BW): + 2 dB/-3 dB
Low Frequency Response
(AC-3 dB)
≤10 Hz (at input BNC)
Noise
STDEV≤0.6 div (≥ 5 mV/div)
STDEV≤0.7 div (2 mV/div)
DC Gain Accuracy
≤ ±3.0%: 5 mV/div ~10 V/div
≤±4.0%: ≤2 mV/div
DC Measurement Accuracy
± [3%× (|reading|+|offset|) +1%×|offset| +0.2 div+2 mV] , ≤100 mV/div
±[3%× (|reading|+|offset|) +1%×|offset| +0.2 div+100 mV] , >100 mV/div
Rise time
Typical 2.3 ns (SDS1152 CML+)
Typical 3.5 ns (SDS1102 CML+)
Typical 5.0 ns (SDS1072 CML+)
Typical 7.0 ns (SDS1052 DL+)
Overshoot (500 ps Pulse) <10%
Trigger System
Trigger Mode Auto, Normal, Single
Trigger Level Range
Internal: ±6 divisions from center of screen
EXT: ±1.2 V
EXT/5: ±6 V
Hold off Range 100 ns ~ 1.5 s
Trigger Coupling AC, DC, LF Rej, HF Rej
Trigger Sensitivity
1 Divisions: DC-10 MHz
1.5 Divisions: 10 MHz - Max BW
Trigger Displacement
Pre-trigger: Memory depth/ (2*sampling)
Delay Trigger: 260 div
Edge Trigger
Slope Rising, Falling, Rising & Falling
Source CH1/CH2/EXT/(EXT/5)/AC Line
Slope Trigger
Slope Rising, Falling
Limit Range <, >, =
Source CH1/CH2
Time Range 20 ns ~ 10 s
Pulse Trigger
Polarity +wid , -wid
Limit Range <, >, =
Source CH1/CH2
Pulse Range 20 ns - 10 s
Video Trigger
Signal Standard NTSC, PAL/Secam
Source CH1/CH2
Trigger condition odd eld, even eld, all lines, line num