Data Sheet

SDM3045X Digital Multimeter
WWW.SIGLENT.COM
Frequency and Period Characteristic
Accuracy± (% of Reading + count)
[1]
Capacitance Characteristic
Accuracy± (% of Reading + count)
[1]
Temperature Characteristic
Accuracy± (% of Reading)
[1]
Function Range Frequency Range Resolution
Accuracy
(one year;
23
±5
)
Frequency /Period 600 mV to 750 V
[2]
20 Hz – 2 KHz 0.01+3
2 KHz – 20 KHz 0.01+2
20 KHz – 200 KHz 0.01+2
200 KHz –500 KHz 0.01+2
Remarks:
[1] Specications are for 0.5 Hour warm-up.
[2]Except for special marks, the AC input voltage is 5% to 110% of range when <100 kHz and 10% to 110% of range when >100 kHz. 750 V range is limited to
750 Vrms.The accuracy is 10 times % of Reading when the measurement range of AC voltage is in 600 mV range.
Function Range
[2]
Max Testing Current Resolution
Accuracy
(one year;
23
±5
)
Capacitance
2 nF 10 μA 0.001 nF 3+10
20 nF 10 μA 0.01 nF 1+10
200 nF 100 μA 0.1 nF 1+9
2 μF 100 μA 0.001 μF 1+10
20 μF 1 mA 0.01 μF 1+10
200 μF 1 mA 0.1 μF 1+9
10000 μF 1 mA 1 μF 2+50
Remarks:
[1] Specications are for 0.5 Hour warm-up and “REF” operation. Using of non-lm capacitor may generate additional errors.
[2] Specications are for from 1% to 110% on 2 nF range and ranges from 10% to 110% on other ranges.
Function
Probe
Type
Probe
Model
Working
Temperature Range
Accuracy
(one year;
23
±5
)
Temperature
coefcient
0
~ 18
28
~ 5 0
Temperature
RTD
[2]
α=0.00385 -200
~ 660
0.16
0.09
TC
[3]
B 0
~ 1820
0.76
0.14
E -270
~ 1000
0.5
0.02
J -210
~ 1200
0.5
0.02
K -270
~ 1370
0.5
0.03
N -270
~ 1300
0.5
0.04
R -50
~ 1760
0.5
0.09
S -50
~ 1760
0.6
0.11
T -270
~ 400
0.5
0.03
Remarks:
[1] Specications are for 0.5 Hour warm-up, not include probe error.
[2] Specications are for 4-wire measure or 2-wire measure under “REF” operation.
[3] Built-in cold terminal compensation for thermocouple, accuracy is ±2
.