Data Sheet

WWW.SIGLENT.COM
SDM3045X Digital Multimeter
AC Characteristic Accuracy± ( % of Reading + count)
1]
Function Range
[2]
Frequency Range Resolution
Accuracy
(one year;
23
±5
)
True-RMS
AC Voltage
[3]
600 mV
20 Hz – 45 Hz 0.01 mV 2.0 + 20
45 Hz – 100 Hz 0.01 mV 0.6 +10
100 Hz – 20 KHz 0.01 mV 0.2 + 10
20 KHz – 50 KHz 0.01 mV 1.0 + 10
50 KHz –100 KHz 0.01 mV 3.0 + 10
6 V
20 Hz – 45 Hz 0.0001 V 2.0 + 20
45 Hz – 100 Hz 0.0001 V 0.6 + 10
100 Hz – 20 KHz 0.0001 V 0.2 + 10
20 KHz – 50 KHz 0.0001 V 1.0 + 10
50 KHz –100 KHz 0.0001 V 3.0 + 10
60 V
20 Hz – 45 Hz 0.001 V 2.0 + 20
45 Hz – 100 Hz 0.001 V 0.6 +10
100 Hz – 20 KHz 0.001 V 0.2 + 10
20 KHz – 50 KHz 0.001 V 1.0 + 10
50 KHz –100 KHz 0.001 V 3.0 + 10
600 V
20 Hz – 45 Hz 0.01 V 2.0 + 20
45 Hz – 100 Hz 0.01 V 0.6 + 10
100 Hz – 20 KHz 0.01 V 0.2 + 10
20 KHz – 50 KHz 0.01 V 1.0 + 10
50 KHz –100 KHz 0.01 V 3.0 + 10
750 V
20 Hz – 45 Hz 0.01 V 2.0 + 20
45 Hz – 100 Hz
[4]
0.01 V 0.6 + 10
100 Hz – 20 KHz 0.01 V 0.2 + 10
20 KHz – 50 KHz 0.01 V 1.0 + 10
50 KHz –100 KHz 0.01 V 3.0 + 10
True-RMS
AC Current
[5]
60 mA
20 Hz – 45 Hz 0.001 mA 2.0 + 20
45 Hz – 2 KHz 0.001 mA 0.5 + 20
2 KHz – 10 KHz 0.001 mA 2.5 + 30
600 mA
20 Hz – 45 Hz 0.01 mA 2.0 + 20
45 Hz – 2 KHz 0.01 mA 0.5 + 20
2 KHz – 10 KHz 0.01 mA 2.5 + 30
6 A
20 Hz – 45 Hz 0.0001 A 2.0 + 20
45 Hz – 2 KHz 0.0001 A 0.5 + 20
2 KHz – 10 KHz 0.0001 A 2.5 + 20
10 A
[6]
20 Hz – 45 Hz 0.001 A 1.5 + 15
45 Hz – 2 KHz 0.001 A 0.5 + 15
2 KHz – 10 KHz 0.001 A 2.5 + 25
Additional wave crest factor error ( not Sine )
[7]
Wave crest coefcient Error (% Range)
1-2 0.05
2-3 0.3
Remarks:
[1] Specications are for 0.5 Hour warm-up, “Slow” measurement rate and calibration temperature 18
~ 28
.
[2] 10% over range on all ranges except for DCV 1000 V, ACV 750 V, DCI 10 A and ACI 10 A.
[3] Specications are for amplitude of sine wave input > 5% of range. For inputs from 1% to 5% of range and <50 kHz, add 0.1% of range extra error. For 50 kHz
to 100 kHz, add 0.1% of range extra error.
[4] Plus 0.025 V of error per 1 V after the rst ±400 VAC.
[5] Specications are for sine wave input > 5% of range. 0.1% errors will be added when the range of input sine wave is 1% to 5%.
[6] 30 seconds OFF after 30 seconds ON is recommend for the continuous current that higher than DC 7 A or AC RMS 7 A.
[7] For inputs Frequency Range < 100 Hz