Data Sheet

SDM3045X Digital Multimeter
WWW.SIGLENT.COM
Specications
DC Characteristic Accuracy± ( % of Reading + count)
[1]
Function Range
[2]
Test current or Load
voltage
Resolution
Accuracy
(one year;
23
±5
)
DC Voltage
600 mV 0.01 mV 0.01+ 5
6 V 0.0001 V 0.01+ 6
60 V 0.001 V 0.02+ 4
600 V 0.01 V 0.02+ 6
1000 V
[4]
0.1 V 0.02+ 6
DC Current
600 μA < 33 mV 0.01 μA 0.05+ 3
6 mA < 330 mV 0.0001 mA 0.05+ 3
60 mA < 0.05 V 0.001 mA 0.05+ 3
600 mA < 0.5 V 0.01 mA 0.12+ 6
6 A < 0.33 V 0.0001 A 0.20+ 5
10 A
[5]
< 0.6 V 0.001 A 0.25+ 4
Resistance
[3]
600 Ω 1 mA 0.01 Ω 0.04+ 5
6 KΩ 1 00 μA 0.0001 KΩ 0.02+ 5
60 KΩ 10 μA 0.001 KΩ 0.02+ 5
600 KΩ 1 μA 0.01 KΩ 0.04+ 5
6 MΩ 200 nA 0.0001 MΩ 0.12+ 3
60 MΩ 200 nA || 10 MΩ 0.001 MΩ 0.85+ 3
100 MΩ 200 nA || 10 MΩ 0.01 MΩ 1.75+ 3
Diode Test
[6]
0~2 V 1 mA 0.0001 V 0.05+ 3
2~4 V 1 mA 0.0001 V 0.35+ 3
Continuity Test 2000 Ω 1 mA 0.1 Ω 0.05+ 3
Remarks:
[1] Specications are for 0.5 Hour warm-up, “Slow” measurement rate and calibration temperature 18
~ 28
.
[2] 10% over range on all ranges except for DCV 1000 V, ACV 750 V, DCI 10 A and ACI 10 A.
[3] Specications are for 4-wire measure or 2-wire measure under “REF” operation. ±0.2 Ω of extra errors will be generated if perform 2-wire measure without “REF”
operation.
[4] Plus 0.02 mV of error per 1 V after the rst ±500 VDC.
[5] 30 seconds OFF after 30 seconds ON is recommend foe the continuous current that higher than DC 7 A or AC RMS 7 A.
[6] Accuracy specications are only for voltage measuring at input terminal. The typical value of current under measure is 1 mA. Voltage drop at diode junction may
vary with current supply. Adjustable voltage range : 0~ 4 V.