Quick Start Guide
SIGLENT Chapter 3 System Utility Function
SDL1000X User Manual 88
2. Turn on the “TMon” tab to enable the I_Protect. Then set the protection
current and delay time, as shown in Figure 3-17.
Figure 3-17 I_Protect tab
3. Connect the load to the DC power under test when the power is OFF.
4. Set a sink current in CC mode of the load.
5. Turn on the power output and the load input. Delay a period of time, the
overcurrent protection of the load will occur.
6. Close all instruments and finish the test.
P_Protect
SDL1000X electronic load can simulate overpower protection by the
software of the load. Turn on the load input, if the sink power of the load is
higher than the protection power, the load will turn off input automatically and
pop-up over power protection message after a delay period.
Operating Steps
1. Press Shift and CP key to enter system utility function interface. Then
enter “Config” tab and select “P_Protect” tab to enter P_Protect function.
2. Turn on the “TMon” tab to enable the P_Protect. Then set the protection
power value and delay time, as shown in Figure 3-17.