Quick Start Guide

Chapter 3 System Utility Function SIGLENT
SDL1000X User Manual 87
Figure 3-16 SLTM tab
3. Input the low voltage and high voltage, then turn on the TMon
4. Connect the load to the DC power under test. Set the power output higher
than the V_High. At this time, the power is OFF.
5. Set a sink current of the load in CC mode, then turn on the load input.
6. Turn on the power under test, The T_Rise is displayed quickly on the LCD
screen which is the voltage rise time.
7. Turn off the power under test, The T_Fail on the LCD screen is the falling
time.
8. Close all instruments and finish the test.
Limit
I_Protect
SDL1000X electronic load can overcurrent protection by the software of
the load. Turn on the load input, if the sink current of the load is higher than the
protection current, the load will turn off input automatically and pop-up an
overcurrent protection message after a delay period.
Operating Steps
1. Press Shift and CP key to enter system utility function interface. Then
enter Config tab and select I_Protect tab to enter I_Protect function.