Quick Start Guide

SIGLENT Chapter 3 System Utility Function
SDL1000X User Manual 86
Figure 3-15 Waveform display interface of ExtV mode
ExtSwitch
The load input switches can be controlled by the external TTL electrical
level. During external input control, the [On/Off] key will become invalid
and the load input switches can only be controlled by the external TTL
electrical level. The load input will be switched on in case of low level
external input and meet a falling edge; and the load input will be switched
off in case of high level external input.
IMON
The 0-10 V analog quantity output signal of current monitoring output
terminal represents input current to which the terminal belongs from 0 to
full range. The input current is proportional to the voltage on the terminal.
An external voltmeter or oscilloscope can be connected to display input
current change.
VMON
The 0-10 V analog quantity output signal of voltage monitoring output
terminal represents input voltage to which the terminal belongs from 0 to
full range. The sink voltage is proportional to the voltage on the terminal.
An external voltmeter or oscilloscope can be connected to display input
current change.
SLMT
SDL1000X electronic load can measure the rise and fall time of the voltage
in CC, CV, CR and CP mode. This function can simply simulate the rise
and fall rate of the power voltage.
Operating Steps
1. Press Shift and CP key to enter system utility function interface. Then
enter Config tab and select SLMT tab to enter SLMT function.
2. Turn on the TMon tab, The voltage rise time T_Rise and falling time
T_Fall will be displayed on LCD screen, as shown in figure 3-16.