Figure 3-11 V - I characteristic curve of Von Latch
Set Trigger
Trigger operation is mainly applied to the Dynamic, List and Program
test functions. The load supports three modes: Manual, Ext and Bus.
Manual
When the trigger source is set to Manual mode, press the trig tab
in dynamic, , List, or Program mode will generate a trigger signal.
Bus
When the trigger source is set to Bus mode, the load will execute a
trigger operation when it receives the remote command (such as
*TRG) from USB, LAN, RS232 and GPIB interface.
Manual
The Ext monitor terminals on the rear panel of the load can receive
external trigger signals. When the trigger source is set to be Ext mode and a
pulse(0~5V TTL) which have falling edge plus to the terminal, the load will
trigger signal.
Set Aver
In “Config” tab, when user select “Aver” can set the average point number
of the read-back current and voltage in the load. The range is 6~14. The