Figure 2-52 Sense connection
Operating steps
1. As shown in the above figure, connect the load to the DUT. Ensure
that the polarity of the connections is correct.
2. Press Shift and CP key to enter the system utility function
interface. Press Config and Sense to enable (disable) the Sense
function.
Note: The Sense terminals must be connected to the output
terminals of the DUT. If not, then the load cannot accurately
measure the voltage of the terminals in any function.
Tip: In case where the DUT outputs large currents, the load leads
should be as short as possible. Twisting the sense leads can also
assist in minimizing measurement noise.
External Trigger Function
The SDL features an external trigger which allows sequencing and control
to be performed using another instrument or trigger source. To start, set
the trigger source to External by pressing Shift > Utiity > Config > Page 2/2
and set Trig to Ext. The trigger signal input and ground connection are
shown on the rear panel. The trigger input signal should be 0~5V TTL level