Quick Start Guide

Chapter 2 Function and Features SIGLENT
SDL1000X User Manual 53
(10) Set input value
Set running current value in CC mode. The default unit is A
Set running voltage value in CV mode. The default unit is V
Set running power value in CP mode. The default unit is W
Set running resistance value in CR mode. The default unit is Ω
(11) Set Rrange
Resistance range includes Low, Middle, High and Upper.
(12) Set LED operating voltage (Vo)
Set the operating voltage value in LED mode. The default unit is V
(13) Set LED operating current (Io)
Set the operating current value in LED mode. The default unit is A
(14) Set LED Coeff (Rco)
Users can change the operating voltage and resistance by
changing Rco which has a set range from 0~1.
3. Turn on the input channel
Press On/Off to turn on the channel input. At this time, the actual
discharge voltage, current, time and capacity will be displayed on the
main interface. The parameter status can be displayed in real time in
parameter preview area.
NOTE: The test operation depends on the trigger setting. If you have
selected manual trigger, you can press Trigger to being the
programmed test at this time.
Warning
To avoid electric shock, ensure that the DUT is connected to the
input terminals of the load properly before you turn on the channel
input.
4. View waveform
Press Display key to enter the waveform display interface, as shown in
figure 2-44. Here, you can observe the change of the
current/voltage/resistance/power over time. Press the Display key
again to exit the waveform display interface and return to the main
interface of the Program test mode.