SIGLENT Chapter 2 Function and Features
SDL1000X User Manual 52
LED.
(2) Set the current and voltage ranges:
Current range: 5 A or 30 A
Voltage range: 36 V or 150 V
(3) Set pause Status
User can pause auto test if need. Simply highlight/select the cell
labeled paus using the keypad arrows on the front panel and rotate
the SDL control knob to enable change the entry value. Then the
load will pause at this step when running this auto test step.Auto test
can recover after receiving the next trigger.
(4) Set short Status
User can enter a short circuit step if needed. Similar to pause, use
the keypad arrows to select the Short cell and rotate the knob to
enable it. Then the load will short the circuit at this step when
running an auto test.
(5) Set Loading Time(Ton)
This is the time duration of each step and its range is 0.01~100
seconds (s).
(6) Set unLoading Time(Toff)
The time interval in adjacent step and its setting range is 0.01~100S
(7) Set Delay Time(Tdly)
The time is between begin test to sample the current or voltage
value of the load. Its setting range is 0.01~Ton s.
(8) Set minimum
The minimum sink value. Units depend on the current mode in use.
In CC mode, its default unit is V
In CV mode, its default unit is A
In CP mode, its default unit is W
In CR mode, its default unit is Ω
(9) Set maximum
The maximum sink value Units depend on the current mode in use.
In CC mode, its default unit is V
In CV mode, its default unit is A
In CP mode, its default unit is W
In CR mode, its default unit is Ω