Quick Start Guide
Chapter 2 Function and Features SIGLENT
SDL1000X User Manual 51
can also change the test function type (CC, CV, CP, CR, LED, etc..) at every
step. Up-to ten test files can be saved with each test file having a maximum of
fifty steps each. The file extension defined for an Autotest file is “.prog”.
Operating Steps
1. Connect device
Power on the instrument and connect the DUT and the channel input
terminals of the electronic load, as shown in Figure 2-2.
2. Set running parameters
Press the Shift + CV key on the front panel , select “Program” then
enter Program test mode, as shown in 2-45.
Figure 2-45 Program Test Function
The parameters for the program test mode include Step, Storage, Trig,
Result.
Set Step
Set the number of steps you wish to run in this specific program.
Set parameter of list
The parameter list of Program mode includes mode, current range
(Irange), voltage range (Vrange), pause (paus), Short, time on (Ton),
time off (Toff), delay (Tdly), min, max, Set, resistance range (Rrange),
Vo, Io, Rco. The arrow in the right of parameter list indicates that the
current interface can not display parameters of all steps. At this time,
user can edit all parameters of each steps by using direction key after
move the focus to the list.
(1) Set operating mode
There are five modes that can be selected: CC, CV, CP, CR, or