SIGLENT Chapter 2 Function and Features
SDL1000X User Manual 34
Figure 2-26 CC Tog mode page 3
The parameters for the pulse operation mode include Function, Mode,
range, A_ Level, B_ Level, rising slew rate, falling slew rate and trigger
selection. The interface menu can be divided into three pages.
Set range
Current range: 5 A or 30 A
Voltage range: 36 V or 150 V
Set A_level
The sink current toggles between a high value and a low value in Pul
mode. A_Level indicates the low value. The default unit for A_Level is
Ampere (A).
Set B_level
The sink current toggles between a high value and a low value in Pul
mode. B_Level indicates the high value. The default unit for B_Level is
Ampere (A).
Slew_RIS and Slew_FAIL
Slew_RIS and Slew_FAIL: The rate of change to a new level from the
current sink value of the load. Its unit is A/us and the setting range is
0.001 ~ 2.5 A/us.
3. Enable trigger
Press Shift + CP key to enter Utility interface .After pressing “Config”
key the trigger source can be set in page 2 of the menu. There are
three trigger types:Manual/Ext/Bus. Manual displays a trigger softkey