Quick Start Guide
SIGLENT Chapter 2 Function and Features
SDL1000X User Manual 28
Set B_level
The sink current toggles between a high value and a low value in
continuous mode. The B_Level indicates a high value. The default unit
for B_Level is Ampere (A).
Set width
A_width/B_width: The time during which the sink current stays at Level
A. Default units are s or ms and the setting range is 0.020 ms ~ 999 s.
Slew_RIS and Slew_FAL
Slew_RIS and Slew_FAL: The set rate of change to a new level from
the current value. Its unit is A/us and setting range is 0.001 ~ 2.5 A/us.
3. Enable trigger
Press Shift + CP key to enter the Utility interface. After pressing the
“Config” key, the trigger source can be set in page 2 of the selection
menu which includes Manual/Ext/Bus.Manual displays a trigger
softkey on page 3 of the Dynamic test. Pressing this key will trigger the
step. Ext sets the trigger type to external, which looks for a valid trigger
on the rear panel Ext Trig input. Bus is a trigger sent over the remote
control bus. See “Set trigger” in Section 3 for more details. Press
On/Off to turn on the channel input. At this time, the actual input
voltage, current, resistance and power will be displayed on the main
interface. The sink current will toggle continuously between the set A
level and B levels.
Tip:
The sink current may maintain a constant value in the main interface
when the width is set to a very small value because the load is
switching too quickly between A_level and B_level set points. Use
the waveform display function to see the waveform more clearly.
4. View waveform display
Press Display to enter waveform display interface, shown in figure
2-18. The waveform will display the current curve when ”I” is selected
in CC Cont mode. Press Display again to exit the waveform display
interface and return to the main interface of CC Cont mode.