SDL1000X User Manual 26
Continuous mode
In continuous operation, when you enable the dynamic test operation, the
load will continuously switch between Level A and Level B, as shown in Figure
2-14.
Figure 2-14 Transient CC Continuous Mode (Cont)
Take CC mode as example (other modes function in a similar fashion)
Operating Steps
1. Connect device
Power on instruments Connect the DUT and the channel input
terminals of the electronic load, as shown in Figure 2-2.
2. Set running parameters
Press the Shift + CC key on the front panel to enter transient test
operation. Continuous mode (CC) requires setting the proper slew
rates for all Level steps, as shown in 2-15, 2-16, 2-17.
Figure 2-15 CC Cont mode page 1