While making a connection, the positive polarity of the load should
be connected to the (+) terminal of the channel output, and the
negative polarity of the load to the (-) terminal of the channel
output. Incorrect polarity may cause damage to the instrument or
the DUT.
2. Press CP to enter the main interface of CP mode, as shown in
Figure 2-12.
Figure 2-12 CP Mode Main Interface
3. Set CP mode current range (5 A or 30 A) and voltage range (36 V
or 150 V).
4. Set power value.
5. Press On/Off to turn on the channel input. At this time, the actual
input voltage, current, resistance and power will be displayed on
the main interface.
Note: The load begin to sink power only if the input voltage is greater
than the conduction voltage of the system (default value is 0 V)
Warning
To avoid electric shock, ensure that the DUT is connected to the
input terminals of the load properly before you turn on the channel
input.
6. Press the Display key to enter the waveform display interface, as
shown in Figure 2-13. By default, the power waveform is displayed.
When the input voltage changes, the load will sink a constant
power. Press the Display key again to exit the waveform display