Quick Start Guide

Chapter 2 Function and Features SIGLENT
SDL1000X User Manual 17
Figure 2-2
CAUTION
While making a connection, the positive polarity of the load should
be connected to the (+) terminal of the channel output, and the
negative polarity of the load to the (-) terminal of the channel
output. Incorrect polarity may cause damage to the instrument or
the DUT.
2. Press CC to enter the main interface of CC mode, as shown in
Figure 2-3.
3.
Figure 2-3 CC Mode Main Interface
4. Set CC mode current range (5 A or 30 A) and voltage mode (36 V
or 150 V)
Note: The lower ranges provide better resolution and accuracy at
low current settings.
5. Set the current input value. This is the amount of current that the
load will attempt to draw when the output is enabled.
6. Set the rising slew rate and the falling slew rate in CC mode. The
default unit for the slew rate is A/us.
7. Press On/Off to turn on the channel input. At this time, the actual
input voltage, current, resistance and power will be displayed on
the main interface.
Note: The load will begin to sink current only if the input voltage is
greater than the conduction voltage of the system (default
value is 0 V).