User Manual SDL1000X Programmable DC Electronic Load UM0801X-C01A 2019 SIGLENT TECHNOLOGIES CO.
SIGLENT Copyright and Statement Copyright SIGLENT TECHNOLOGIES CO., LTD. All rights reserved. Trademark Information SIGLENT is registered trademark of SIGLENT TECHNOLOGIES CO., LTD. Statement ● SIGLENT products are protected by patent laws in and outside of the P.R. China. ● SIGLENT reserves the right to change the specifications and price. ● Information in this publication replaces all previous corresponding published material.
SIGLENT General Safety Summary Please review the following safety precautions carefully to avoid personal injury or damage to this product or any product connected to it. To prevent potential danger, please use the instrument as specified. Use the proper power cord Only use the the power cord designed for use with the instrument and authorized for use with in the local country of interest. Power supply AC Input Voltages: 110 V/220 V ± 10%, 50/60Hz.
SIGLENT Relative humidity: < 80% Altitude: < 2000 m Temperature: 0℃ to 40℃ Electrostatic Prevention Operate in an area that minimizes the potential for electrostatic discharge . Always ground both the internal and external conductors of cables to release static before connecting. Do not operate in an explosive atmosphere To avoid personal injury or damage to instrument, please do not operate in an explosive atmosphere.
SIGLENT Safety Terms and Symbols Important terms that may appear on the product and in this manual: DANGER: Indicates direct injury or hazard that may happen. WARNING: Indicates potential injury or hazard that may happen. CAUTION: Indicates potential damage to the instrument or other property that may happen.
SIGLENT SDL1000X Brief Introduction The SDL1000X series Programmable DC Electronic Load has a 3.5 inch TFT-LCD display , a user-friendly interface and superb performance specifications. Two versions are available: The SDL1020X features an input range of 150 V/30 A with a total power of 200 W. The SDL1030X has an input range of with a total power of 300 W. All models have measurement resolutions of 0.1 mV/0.1 mA,Adjustable current rise time of 0.001 A/s ~ 2.
SIGLENT Min. readback resolution: 0.1 mV, 0.1 mA Measuring speed of voltage and current: Up to 500 kHz 100 step list function Over-current protection, over-power protection , battery, short circuit and CR-LED test functions 4-wire SENSE compensation mode External voltage and current control function Voltage and current monitoring via 0-10 V analog output 3.
SIGLENT Content Copyright and Statement II General Safety Summary III Safety Terms and Symbols V SDL1000X Brief Introduction VI Chapter 1 Start Guide 1 General Inspection 2 The Front Panel 3 The Rear Panel 6 Connect power8 User interface 10 Fuse Replacement 13 Chapter 2 Function and Features 14 Local/Remote Operation Mode 15 Local Operation Mode 15 Remote Operation Mode 15 Static Operation Mode 16 Constant Current (CC) Mode 16 Constant Voltage (CV) Mode 18 Constant Resistance (CR) Mode 21 Constant Power (C
SIGLENT Overcurrent protection (OCP) 67 Overpower protection (OPP) 67 Over-temperature Protection (OTP) 68 Input Reverse Polarity Protection (RPP) 68 Chapter 3 System Utility Function 69 System 69 1.System info 70 2.Interface71 2.Sound 75 3.Language 76 4.Factory Setting 76 5.Upgrade 77 6.
Chapter 1 Start Guide SIGLENT Chapter 1 Start Guide In this chapter, we introduce the front panel and display interface of the SDL1000X, and also tips for how to check and operate the digital load for the first time.
SIGLENT Chapter 1 Start Guide General Inspection Please check the instrument according to the following steps. 1. Inspect the shipping container. Keep the shipping container or cushioning material until the contents of the shipment have been completely checked and the instrument has passed both electrical and mechanical tests. The consigner or carrier will be responsible for damage to the instrument resulting from shipment. SIGLENT will not provide free maintenance or replacement. 2.
Chapter 1 Start Guide SIGLENT The Front Panel 3 1 2 5 4 6 7 Figure1: The front panel of the SDL1000X 1. LCD The load has a 3.5 inch TFT-LCD display which can display system parameter settings, system output state, wave forms, menu options, prompt messages, etc. 2. Knob When setting parameters, rotate the knob to increase or decrease the value of the digit at the cursor.
SIGLENT Chapter 1 Start Guide 3. Function button and power key Press down the button to enter the constant current mode. Enter one of the dynamic modes (DYN mode) by pressing the shift button at the same time. Press down the button to enter constant voltage (CV) mode. Enter auxillary (AUX) mode by pressing the shift button at the same time. Press down the button to enter the constant power (CP) mode. Enter Utility mode by pressing the shift button at the same time.
Chapter 1 Start Guide SIGLENT cursor to select the setting parameter. 0~9 Input the number zero to nine using the keypad. Dot/Period Enter Softkey functions are define by the adjacent on-screen label 4. Input Terminal Physical input connections to the external circuit and voltage. 5. USB interface Interface that can be insert USB device and support FAT32 file system format. 6. Power key Turn on or off the instrument. 7. Function key Used to choose different functions.
SIGLENT Chapter 1 Start Guide The Rear Panel 1 9 10 5 2 3 4 12 11 8 6 7 Figure 2: The rear panel of the SDL1000X 1. Warning message Warnings about grounding the instrument and other important information. 2. AC input voltage description The frequency and voltage of the AC power supply must match the specification of the fuse. 3. AC power socket The power plug socket for AC input power. 4.
Chapter 1 Start Guide SIGLENT The specified fuse must be rated for the input voltage (Please refer to the “ AC input voltage description”) 5. AC line power selection switch AC Input Voltage: 110/220 V 6. LAN interface Connect to the local network (LAN) using a standard RJ45 interface. 7. USB device Connect the instrument (as a controlled device) to an external computer/controller via USB. 8. GPIB interface Connect to a controlling computer GPIB card through the IEEE488 bus. 9.
SIGLENT Chapter 1 Start Guide Connect power The SDL1000X supports a variety of AC line power input values. For each line voltage, the rear panel voltage selector settings are different, as shown in table 1 below. Table 1: AC input line power specifications AC power input Voltage selector configure 110 VAC ± 10%, 50 ~ 60 Hz 220 VAC ± 10%, 50 ~ 60 Hz Please connect the power carefully follow the steps below: 1.
Chapter 1 Start Guide SIGLENT 4. Connect the power Connect the instrument to AC power supply using the power cord provided in the accessories. Then press the button to turn on the electronic load. WARNING Before turning the instrument on, please disconnect the power supply and set the voltage selector to the appropriate value. WARNING To avoid electric shock, make sure that the instrument is correctly grounded.
SIGLENT Chapter 1 Start Guide User interface Figure 3: The user interface of the SDL1000X 1. Channel output mode 2. Channel output state 3. Short state 4. Remote sense mode 5. LAN connection icon 6. USB connection icon 7. Remote mode 8.
Chapter 1 Start Guide 9. Setting value 10. Measured output values 11.
SIGLENT Chapter 1 Start Guide To Power on the instrument After the instrument is connected to the power source, press the Power key at the left bottom of the front panel to power on the instrument. When the instrument is turned on, it will undergo a self-test. If the instrument passes the self-test, the welcome interface is displayed. If the instrument fails, the self-test failure information will be displayed. If you encounter any problems or failures, please contact your nearst SIGLENT support office.
Chapter 1 Start Guide SIGLENT Fuse Replacement The specifications of the fuse are relative to the actual input line voltage shown in the table below. You also can refer to the rear panel “input power requirement”. Input voltage Fuse specification 110 VAC T315mA 220 VAC T315mA To replace the fuse, please follow the steps below: 1. Turn off the instrument and remove the power cord. 2. Insert a small straight screwdriver into the slot at the power socket and gently pry out the fuse seat. 3.
SIGLENT Chapter 2 Function and Features Chapter 2 Function and Features Contents in this chapter: Local/Remote Operation Mode Static Operation Mode Transient Test Function OCPT Test Function OPPT Test Function Auto Test Function LED Test Function Waveform Display Function Store and Recall Rear Panel Terminal functions Short monitor function Protective function SDL1000X User Manual 14
Chapter 2 Function and Features SIGLENT Local/Remote Operation Mode The load provides two operation modes: Local and remote. Local Operation Mode After you power on the instrument, it enters the local operation mode by default. In the local operation mode, all the keys on the front panel are available for you to use. Remote Operation Mode In the remote operation mode, you can send programming commands from a controller (computer) via any one of the interfaces (GPIB, USB, RS232, or LAN).
SIGLENT Chapter 2 Function and Features Static Operation Mode The static operation modes include the following 4 modes: Constant Current (CC) Mode Constant Voltage (CV) Mode Constant Resistance (CR) Mode Constant Power (CP) Mode Constant Current (CC) Mode In CC mode, the electronic load will sink a current in accordance with the programmed value regardless of the input voltage, as shown in Figure 2-1.
Chapter 2 Function and Features SIGLENT Figure 2-2 CAUTION While making a connection, the positive polarity of the load should be connected to the (+) terminal of the channel output, and the negative polarity of the load to the (-) terminal of the channel output. Incorrect polarity may cause damage to the instrument or the DUT. 2. Press CC to enter the main interface of CC mode, as shown in Figure 2-3. 3. Figure 2-3 CC Mode Main Interface 4.
SIGLENT Chapter 2 Function and Features Warning To avoid electric shock, ensure that the DUT is connected to the input terminals of the load properly before you turn on the channel input. 8. Press the Display key to enter the waveform display interface, as shown in Figure 2-4. By default, the current waveform is displayed. When the input voltage changes, the load will sink a constant current. Press the Display key again to exit the waveform display interface and return to the main interface of CC mode.
Chapter 2 Function and Features SIGLENT Constant voltage mode Figure 2-5 Voltage-Current Relationship Schema under CV Mode Operating Steps 1. Turn off the instrument, as shown in Figure 2-2, connect the DUT and the channel input terminals on the front panel of the load. CAUTION While making a connection, the positive polarity of the load should be connected to the (+) terminal of the channel output, and the negative polarity of the load to the (-) terminal of the channel output.
SIGLENT Chapter 2 Function and Features or 150 V) Note: The low range provides better resolution and accuracy at low voltage settings. 4. Set voltage 5. Press On/Off to turn on the channel input. At this time, the actual input voltage, current, resistance and power will be displayed on the main interface. Note: The load will sink current only if the input voltage is larger than the conduction voltage of the system (default value is 0 V).
Chapter 2 Function and Features SIGLENT Constant Resistance (CR) Mode In CR mode, the electronic load is regard as a constant resistance and will give linear change of current with respect to input voltage changes, as shown in Figure 2-8. Constant resistance mode Figure 2-8 Voltage-Current Relationship Schema under CR Mode Operating Steps 1. Turn off the instrument, as shown in Figure 2-2, connect the DUT and the channel input terminals on the front panel of the load.
SIGLENT Chapter 2 Function and Features Figure 2-9 Main Interface of CR Mode 3. Set CR mode current range (5 A or 30 A), voltage range (36 V or 150 V) and resistance range (Low/Middle/High/Upper). Note: The low range provides better resolution and accuracy at low resistance settings. 4. Set the target resistance value 5. Press On/Off to turn on the channel input. At this time, the actual input voltage, current, resistance and power will be displayed on the main interface.
Chapter 2 Function and Features SIGLENT Figure 2-10 Waveform display interface of CR Mode Constant Power (CP) Mode In CP mode, the electronic load will sink a constant power. If the input voltage rises, the input current draw will be decreased to maintain a constant power sink following the equation (P= V*I), as shown in Figure 2-11. Constant power mode Figure 2-11 Voltage-Current Relationship Schema under CP Mode Operating Steps 1.
SIGLENT Chapter 2 Function and Features While making a connection, the positive polarity of the load should be connected to the (+) terminal of the channel output, and the negative polarity of the load to the (-) terminal of the channel output. Incorrect polarity may cause damage to the instrument or the DUT. 2. Press CP to enter the main interface of CP mode, as shown in Figure 2-12. Figure 2-12 CP Mode Main Interface 3. Set CP mode current range (5 A or 30 A) and voltage range (36 V or 150 V). 4.
Chapter 2 Function and Features SIGLENT interface and return to the main interface of CP mode. Figure 2-13 Waveform display interface of CP Mode Dynamic test function In dynamic test operation, the electronic load can be switched between two parameters based on the set values. This can be useful when testing the dynamic performance of a power supply or source. Press Shift + CC key on the front panel to enter the dynamic test interface.
SIGLENT Chapter 2 Function and Features Continuous mode In continuous operation, when you enable the dynamic test operation, the load will continuously switch between Level A and Level B, as shown in Figure 2-14. Figure 2-14 Transient CC Continuous Mode (Cont) Take CC mode as example (other modes function in a similar fashion) Operating Steps 1. Connect device Power on instruments Connect the DUT and the channel input terminals of the electronic load, as shown in Figure 2-2. 2.
Chapter 2 Function and Features SIGLENT Figure 2-16 CC Cont mode page 2 Figure 2-17 CC Cont mode page 3 The parameters for the continuous operation mode include Function, Mode, range, A _Level, B_Level, rising slew rate, falling slew rate, A_width, B_width and trigger selection. The interface menu can be divided into three pages. Set range Current range: 5 A or 30 A Voltage range: 36 V or 150 V Set A_level The sink current toggles between a high value and a low value in continuous mode.
SIGLENT Chapter 2 Function and Features Set B_level The sink current toggles between a high value and a low value in continuous mode. The B_Level indicates a high value. The default unit for B_Level is Ampere (A). Set width A_width/B_width: The time during which the sink current stays at Level A. Default units are s or ms and the setting range is 0.020 ms ~ 999 s. Slew_RIS and Slew_FAL Slew_RIS and Slew_FAL: The set rate of change to a new level from the current value.
Chapter 2 Function and Features SIGLENT Figure 2-18 Waveform display interface of Cont mode Pulse mode Dynamic test operations using pulse mode configures the load to source the low value (Level A) until a valid trigger is received. At this time, the load settings will change to the B values. The settings will switch back to the A values after maintaining B for the set pulse width time, as shown in Figure 2-19.
SIGLENT Chapter 2 Function and Features the only one mode that requires a set slew rate. Switch to pulse mode by pressing the “Mode” key, as shown in 2-20, 2-21, 2-22.
Chapter 2 Function and Features SIGLENT Figure 2-22 CC pulse mode page 3 The parameters for the pulse operation mode include Function, Mode, range, A_ Level, B_ Level, rising slew rate, falling slew rate, width and trigger selection. The interface menu can be divided into three pages. Set range Current range:5 A or 30 A Voltage range: 36 V or 150 V Set A_level The sink current toggles between a high value and a low value in Pulse mode. The A_Level is designated as the low value.
SIGLENT Chapter 2 Function and Features Press Shift + CP key to enter the Utility interface. After pressing the “Config” key, the trigger source can be set on page 2 of the menu. There are three trigger options available: Manual/Ext/Bus . Manual displays a trigger softkey on page 3 of the Dynamic test. Pressing this key will trigger the step. Ext sets the trigger type to external, which looks for a valid trigger on the rear panel Ext Trig input. Bus is a trigger sent over the remote control bus.
Chapter 2 Function and Features SIGLENT Power on instruments Connect the DUT and the channel input terminals of the electronic load, as shown in Figure 2-2. 2. Set running parameters Press the Shift + CC key on the front panel to enter tansient test operation. CC continuous mode is not only the default mode but also the only one mode that requires setting the slew rate. Switch to Tog mode by press the “Mode” key, as shown in 2-24, 2-25, 2-26.
SIGLENT Chapter 2 Function and Features Figure 2-26 CC Tog mode page 3 The parameters for the pulse operation mode include Function, Mode, range, A_ Level, B_ Level, rising slew rate, falling slew rate and trigger selection. The interface menu can be divided into three pages. Set range Current range: 5 A or 30 A Voltage range: 36 V or 150 V Set A_level The sink current toggles between a high value and a low value in Pul mode. A_Level indicates the low value. The default unit for A_Level is Ampere (A).
Chapter 2 Function and Features SIGLENT on page 3 of the Dynamic test. Pressing this key will trigger the step. Ext sets the trigger type to external, which looks for a valid trigger on the rear panel Ext Trig input. Bus is a trigger sent over the remote control bus. See “Set trigger” in Section 3 for more details. Press On/Off to turn on the channel input. At this time, the actual input voltage, current, resistance and power will be displayed on the main interface.
SIGLENT Chapter 2 Function and Features Figure 2-27 OCPT Test Function Operating Steps 1. Connect device Power on instruments and connect the DUT and the channel input terminals of the electronic load, as shown in Figure 2-2. 2. Set running parameters Press the Shift + CV key on the front panel , select “OCPT” then enter OCPT test operation, as shown in 2-28, 2-29, 2-30.
Chapter 2 Function and Features SIGLENT Figure 2-28 OCPT Test Function Page 1 Figure 2-29 OCPT Test Function Page 2 SDL1000X User Manual 37
SIGLENT Chapter 2 Function and Features Figure 2-30 OCPT Test Function Page 3 The parameters for the OCPT test mode include range, OCP_V, I_Step, I_Start, I_End, I_MIN, I_MAX, Delay. The interface menu can be divided into three pages. Set range Current range: 5 A or 30 A Voltage range: 36 V or 150 V Set OCP_V The protection voltage of the OCPT test function. The default unit for the value is V. Set I_Step The step current in OCPT test function. The default unit for I_Step is Ampere (A).
Chapter 2 Function and Features SIGLENT output. If an OCPT test fails, the interface of the load will pop-up a messagebox such as “Below limit ,test fail”, “Over limit , Test fail”, etc. Then the input channel of the load will be closed automatically. 4. View waveform Press Display key to enter the waveform display interface, as shown in figure 2-31. The sink current should have a stepped increase in the waveform when viewing current (I) data. .
SIGLENT Chapter 2 Function and Features Figure 2-32 OPPT Test Function Operating Steps 1. Connect device Power on instruments and connect the DUT and the channel input terminals of the electronic load, as shown in Figure 2-2. 2. Set running parameters Press the Shift + CV key on the front panel , select “OPPT” then enter OPPT test operation, as shown in 2-33, 2-34, 2-35.
Chapter 2 Function and Features SIGLENT Figure 2-33 OPPT Test Function Page 1 Figure 2-34 OPPT Test Function Page 2 SDL1000X User Manual 41
SIGLENT Chapter 2 Function and Features Figure 2-35 OPPT Test Function Page 3 The parameters for the OPPT test mode include range, OPP_V,P_Step, P_Start, P_End, P_MIN, P_MAX, Delay. The interface menu can be divided into three pages. Set range Current range: 5 A or 30 A Voltage range: 36 V or 150 V Set OPP_V The protection voltage of the OPPT test function. The default unit for the value is Volts (V). Set P_Step The step power for the OPPT test function. The default unit for I_Step is Watts (W).
Chapter 2 Function and Features SIGLENT will pop-up a messagebox that states “Test complete”. Then the load will disable the output. If the OPPT test fails, the interface of the load will pop-up a messagebox such as “Below limit ,test fail”, “Over limit , Test fail”, etc.Then the input channel of the load will closed automatically. 4. View waveform Press Display key to enter waveform display interface, as shown in figure 2-36.
SIGLENT Chapter 2 Function and Features Figure 2-37 OPPT Test Function The Battery test function reflects the reliability of the battery and the remaining battery life. Take CC mode as example (other modes are similar) Operating Steps 1. Connect device Power on the instrument and connect the DUT and the channel input terminals of the electronic load, as shown in Figure 2-2. 2.
Chapter 2 Function and Features SIGLENT Figure 2-38 Battery Test Function Page 1 Figure 2-39 Battery Test Function Page 2 The parameters for the Battery test mode include Function, Range, Current, V_Stop, C_Stop, T_Stop . The interface menu can be divided into two pages. Set Function Choose CC, CR, or CP mode Set range Current range: 5 A or 30 A Voltage range: 36 V or 150 V Set current The discharge current of the Battery test function. The default unit for the value is Ampere (A).
SIGLENT Chapter 2 Function and Features stops discharging automatically. The default unit for C_Stop is milliamp-hour (mAh). Set T_Stop The discharge time in the Battery test function. When the accumulated time is greater than the cut-off time, the load stops discharging automatically. The default unit for T_Stop is seconds (s). 3. Turn on the input channel Press On/Off to turn on the channel input.
Chapter 2 Function and Features SIGLENT List Test Function List test mode enables you to automatically test devices by creating and executing a sequence of test steps using a single function. For example, you could create a list test that contains 10 difference steps using the CC function, or a two step list using CV. You define each step limit, measurement ranges, dwell time, slew rate and trigger mode then save this specific sequence as a list.
SIGLENT Chapter 2 Function and Features Figure 2-42 List Test Function Page 1 Figure 2-43 List Test Function Page 2 The parameters for the List test mode include Function, Range, Step, Count, Storage, and Trig . The interface menu can be divided into two pages.
Chapter 2 Function and Features SIGLENT Set Function Choose CC,, CV, CR, or CP mode Set range Current range: 5 A or 30 A Voltage range: 36 V or 150 V Set Step Here, you set the number of steps that are required for the list. Set Count The Count is the number of times the list will be executed before the end of the test. The Count has a range from 0~255. When Count is 0, the SDL will display “infinite” and will run each step in the list until it is disabled or stopped manually.
SIGLENT Chapter 2 Function and Features parameter preview area. Warning To avoid electric shock, ensure that the DUT is connected to the input terminals of the load properly before you turn on the channel input. 4. View waveform Press Display key to enter the waveform display interface, as shown in figure 2-44. Here, you can observe the measured performance of the input signal. Change the Data label (I, V, P, or R) to view a different parameter.
Chapter 2 Function and Features SIGLENT can also change the test function type (CC, CV, CP, CR, LED, etc..) at every step. Up-to ten test files can be saved with each test file having a maximum of fifty steps each. The file extension defined for an Autotest file is “.prog”. Operating Steps 1. Connect device Power on the instrument and connect the DUT and the channel input terminals of the electronic load, as shown in Figure 2-2. 2.
SIGLENT Chapter 2 Function and Features LED. (2) Set the current and voltage ranges: Current range: 5 A or 30 A Voltage range: 36 V or 150 V (3) Set pause Status User can pause auto test if need. Simply highlight/select the cell labeled paus using the keypad arrows on the front panel and rotate the SDL control knob to enable change the entry value. Then the load will pause at this step when running this auto test step.Auto test can recover after receiving the next trigger.
Chapter 2 Function and Features SIGLENT (10) Set input value Set running current value in CC mode. The default unit is A Set running voltage value in CV mode. The default unit is V Set running power value in CP mode. The default unit is W Set running resistance value in CR mode. The default unit is Ω (11) Set Rrange Resistance range includes Low, Middle, High and Upper. (12) Set LED operating voltage (Vo) Set the operating voltage value in LED mode.
SIGLENT Chapter 2 Function and Features Figure 2-44 Waveform display interface of Program mode 5. Result The auto test complete interface, as shown in Figure 2-45. Result list: Steps, Results and Remarks. Readback parameters: Including readback current, etc. Running step: Indicates the number of running steps. Clear: Remove all data in result list. 6.
Chapter 2 Function and Features SIGLENT Rd: Resistance of operating point A typical V - I characteristic curve of LED lights, as shown in Figure 2-45: Figure 2-45 V - I characteristic curve of LEDs According to the V-I curve, we can get Rd and Vd: 𝑉2 − 𝑉1 Rd = 𝐼2 − 𝐼1 𝑉 =𝑉 − 𝐼 Choose V2, I2, V1, and I1 near the static working point (as shown in the red circle of the curve). Users also can calculate the value of Vd and Rd according to methods below: Vd = V * 0.8 Rd= 0.2 V/I Operating Steps 1.
SIGLENT Chapter 2 Function and Features Figure 2-46 LED Test Function The parameters for the LED test mode include Range, Io, Vo, Rco. Set Range Current range: 5 A or 30 A Voltage range: 36 V or 150 V Set Vo The working voltage at the operating point. Set Io The working current at the operating point. Set Rco Set Rd by the following formula: =(𝑉𝑜/𝐼𝑜 1− 𝑐𝑜) 3. Turn on the input channel Press On/Off to turn on the channel input.
Chapter 2 Function and Features SIGLENT To avoid electric shock, ensure that the DUT is connected to the input terminals of the load properly before you turn on the channel input. 4. View waveform Press Display key to enter waveform display interface, as shown in figure 2-47. Here, you can view data with respect to time. Select a data type to choose the graphed values. Press the Display key again to exit the waveform display interface and return to the main interface of LED test mode.
SIGLENT Chapter 2 Function and Features Figure 2-48 The waveform display in CC mode Operating Steps 1. Set Time The time set range is 4 s ~ 80 h in all modes. Tip When the time that you set is greater than 120 s, it will be automatically displayed in minutes. When greater than 120 min, it will be automatically displayed in hours. 2. Set Data menu items under "Data" include I, U, R, and P. 3.
Chapter 2 Function and Features SIGLENT the data file in ".CSV" format to the USB storage device. This will prompt you to name the file. The load will begin to graph the data as well as write it to the USB memory stick. When you are recording data, the corresponding flag will display in the top of the interface : The flag will disappear after pressing the “Record” key again. 6. Playback Press “PlayBack” to play back the recorded data files.
SIGLENT Chapter 2 Function and Features Note: The load only supports FAT32 USB flash drives. 1. Local (C:) User can save the currently set state files to the local storage area and recall them when necessary. Eight list files and ten program files can be saved to the local drive. 2. USB(D) Disk D is only available when a USB storage device is detected over the USB HOST interface on the front panel.
Chapter 2 Function and Features SIGLENT Figure 2-50 filename edit interface 3. The filename edit box can be input uppercase and lowercase English letters、0~9 natural numbers and special characters “-” or “_” or ”.”, etc. Input character: The character can only be edit in the place where the cursor is located in the filename edit box. Press Enter or “Next Char” key to edit the next character. Character selection area: The white color area is the character selection area in the interface.
SIGLENT Chapter 2 Function and Features Read Operating Steps 1. Press Restore to enter the storage interface, then enter the files catalogue list in List/Program. 2. You can select an arbitrary file in the files catalogue in List/Program, then click “Read” to enter List/Program mode. Copy and Paste This function only operates in copying files from the internal storage (C to an external USB memory device (D:) Operating steps 1.
Chapter 2 Function and Features SIGLENT Function of Terminals on the Rear Panel Sense mode When the DUT outputs large currents, the voltage drop due to the lead and contact resistance of the load leads cannot be ignored. To ensure an accurate measurement for the output voltage of the DUT, the load provides Sense (remote sense) working mode. In this mode, the Sense terminals (as shown in figure 2-51) are directly connected to the output terminals of the DUT.
SIGLENT SDL1000X User Manual 64 Chapter 2 Function and Features
Chapter 2 Function and Features SIGLENT Figure 2-52 Sense connection Operating steps 1. As shown in the above figure, connect the load to the DUT. Ensure that the polarity of the connections is correct. 2. Press Shift and CP key to enter the system utility function interface. Press “Config” and “Sense” to enable (disable) the Sense function. Note: The Sense terminals must be connected to the output terminals of the DUT.
SIGLENT Chapter 2 Function and Features Operating Steps 1. Press Shift and CP key to enter the system utility function menu. Select “Config” and set trigger mode of the trigger source be “Ext”, then press Enter to confirm. 2. When using the external trigger mode, the falling edge of the trigger signal output from the negative and positive trigger terminal can enable trigger function. The trigger can be used in List, Program, and dynamic test mode.
Chapter 2 Function and Features SIGLENT the DUT. Press Shift and Display key to clear overvoltage protection status. Overcurrent protection (OCP) The load supports two kinds of overcurrent protections: Hardware overcurrent protection and software overcurrent protection. Hardware overcurrent protection: The maximum load current of the electronic load is limited to the max current of the existing current range by hardware.
SIGLENT Chapter 2 Function and Features power of the load is greater than the protection power value, the the load input will be disabled and a pop-up overpower message will appear.. To exit OPP protection, press any key on the front panel of the load. Over-temperature Protection (OTP) The load will enter OTP if the temperature of internal power devices exceeds 85℃. If an OTP fault occurs, the load input will be disabled and an OTP warning message will pop-up.
Chapter 3 System Utility Function SIGLENT Chapter 3 System Utility Function System Press Shift and the CP key to enter the system utility function interface. Then press ”system” to enter the system info interface, as shown in figure 3-1 and 3-2. Figure 3-1 System info page 1 Figure 3-2 System info page 2 Table 3-1 Tab Descriptions No.
SIGLENT Chapter 3 System Utility Function 1 System info 2 Interface 3 Sound 4 Language 5 Factory 6 Upgrade 7 BoardTest System info tab. View startup times,software version,hardware version,serial number, and product ID. Communication interface setting tab. Configures parameters for remote communication interfaces (USB/RS232/GPIB/LAN). Sound tab. Enabled/disable sounds. Language tab. Chinese/English Factory tab. Reset the system to the default settings. Upgrade tab.
Chapter 3 System Utility Function SIGLENT 2. Interface Press Shift and the CP key to enter the system utility function interface. Then press the “Interface” key to enter the interface tab of the load which default selection tab is USB, as shown in figure 3-4. Figure 3-4 USB interface The instrument supports USB/RS232/GPIB/LAN communication interfaces. Users can use this remote interface to control the instrument using a computer.
SIGLENT Chapter 3 System Utility Function Use the 9-pin RS232 cable (female-to-female, cross-over) to connect the RS232 interface to the PC, and set the interface parameters (such as baud rate, parity, etc.) that match the PC. Then, user can remotely control the instrument. The RS232 interface is shown in the following figure. For the descriptions of the pins, refer to Table 2-2. Figure 3-5 RS232 Interface Table 3-2 Pin 1 2 3 4 5 6 7 8 9 Pin description of RS232.
Chapter 3 System Utility Function SIGLENT Figue 3-6 RS232 Interface In “RS232” interface, user can set and view relevant interface parameters about RS232. 1. Baud Rate The available baud rates include 4800 9600, 19200, 38400 , 57600 , and 115200 bps. 2. Parity The available parity modes include None, Even, and Odd. 3. Stop Bits The available stop bits include 1 or 2. 4. Data Bits The default data bits is 8. 5. Flow The default flow is None.
SIGLENT Chapter 3 System Utility Function Figure 3-7 GPIB Interface The GPIB Load address range is from 0~30. The default address of GPIB is 1 which is stored in the non-volatile memory and is not influenced by restoration of factory settings. Set the parameters of the LAN interface Before using the LAN interface, connect the load to a computer or the local area network (LAN) where the computer is vissble and ensure sufficient connectivity.
Chapter 3 System Utility Function SIGLENT In this interface, you can view the network connection status and configure the network parameters such as IP address, subnet mask, default gateway and DHCP status. The setting is store in the non-volatile memory and do not influenced by restore factory setting.
SIGLENT Chapter 3 System Utility Function Figure 3-9 Sound tab In the “Sound” menu, you can enable or disable the key sound and tips sound. The load buzzer will sound when you press the key on the front panel or rotate the knob after the key sound is enabled. It will not sound if the key sound is disabled. If the Tips sound is set to ON, the load buzzer will sound when pop-up tips or warning messages. Turn off the Tips sound, and the load buzzer will not sound except during any load protection tips. 3.
Chapter 3 System Utility Function SIGLENT Figure 3-10 Restore factory setting interface When user selects OK, the parameters in every function will restore to defaults. 5. Upgrade The SDL1 can be upgraded with new firmware using an external FAT32 formatted USB memory device. To update the firmware, press Shift and CP key to enter system utility function interface. Then switch to page 2 and press the “Upgrade” key on the front panel.
SIGLENT Chapter 3 System Utility Function switch on the front panel is normal or not. Buzzer: Test whether the buzzer can sound or not. Chip Test: Including the self-test result of the load, EEPROM and ADC.
Chapter 3 System Utility Function SIGLENT Config Turn ON/OFF the Sense Function When the DUT outputs large currents, the Sense terminal can be used to accurately test the voltage across the output terminals to compensate for the voltage drop on the load. Set Vrmt to ON to enable the remote sense test function. Turn ON/OFF SOF Function SOF stands for Stop On Fail. When enabled, the auto test function will stop running immediately when any step fails during an auto test.
SIGLENT Chapter 3 System Utility Function Figure 3-11 V - I characteristic curve of Von Latch Set Trigger Trigger operation is mainly applied to the Dynamic, List and Program test functions. The load supports three modes: Manual, Ext and Bus. Manual When the trigger source is set to Manual mode, press the trig tab in dynamic, , List, or Program mode will generate a trigger signal.
Chapter 3 System Utility Function SIGLENT greater of the average point number, the slower of the read-back current and voltage in the load. Set EXTC (External Interface) The “EXTC” tab includes: (Int, ExtI, ExtV), ExtSwitch, I_M_ON, V_M_ON four tabs. It mainly supply the way user control the external terminals on the rear panel, current and voltage monitor, external control ON/OFF. Int The load will return the mode before enter “EXTC” tab after select the “Int” tab.
SIGLENT Chapter 3 System Utility Function Figure 3-12 The main interface of the ExtI mode The parameters for the ExtI mode include RangI and RangV. Set Range Current range: 5 A or 3 0A Voltage range: 36 V or 150 V Set External Input Voltage The sink current in ExtI mode can be controled by the external input voltage which range is 0~10 V. The default unit of current is A 3. Turn on the input channel Press On/Off to turn on the channel input.
Chapter 3 System Utility Function SIGLENT Note: Once the channel input is turned on, the load will not start to sink the current until the input voltage is greater than the break-over voltage. 4. View waveform Press Display key to enter waveform display interface, as shown in figure 3-13. By means of observe waveform of ExtI mode after set the Data selection be “I”, user can see the trend of the sink current.
SIGLENT Chapter 3 System Utility Function device terminals . 2. Set running parameters Press the Shift + CP key on the front panel to enter the system utility function interface. Select “Config” option and Switch to page 2, then choose “EXTC”. User can enter ExtI interface after press “INT” and choose “ExtV” tab, as shown in 3-14. Figure 3-14 The main interface of the ExtV mode The parameters for the ExtV mode include RangI and RangV.
Chapter 3 System Utility Function SIGLENT 3. Turn on the input channel Press On/Off to turn on the channel input. At this time, the actual sink voltage, current, power, resistance will be displayed on the main interface. Note: Once the channel input is turned on, the load will not start to sink the current until the input voltage is greater than the break-over voltage. 4. View waveform Press Display key to enter waveform display interface, as shown in figure 3-15.
SIGLENT Chapter 3 System Utility Function Figure 3-15 Waveform display interface of ExtV mode ExtSwitch The load input switches can be controlled by the external TTL electrical level. During external input control, the [On/Off] key will become invalid and the load input switches can only be controlled by the external TTL electrical level.
Chapter 3 System Utility Function SIGLENT Figure 3-16 SLTM tab 3. Input the low voltage and high voltage, then turn on the TMon 4. Connect the load to the DC power under test. Set the power output higher than the V_High. At this time, the power is OFF. 5. Set a sink current of the load in CC mode, then turn on the load input. 6. Turn on the power under test, The T_Rise is displayed quickly on the LCD screen which is the voltage rise time. 7.
SIGLENT Chapter 3 System Utility Function 2. Turn on the “TMon” tab to enable the I_Protect. Then set the protection current and delay time, as shown in Figure 3-17. Figure 3-17 I_Protect tab 3. Connect the load to the DC power under test when the power is OFF. 4. Set a sink current in CC mode of the load. 5. Turn on the power output and the load input. Delay a period of time, the overcurrent protection of the load will occur. 6. Close all instruments and finish the test.
Chapter 3 System Utility Function SIGLENT Figure 3-17 P_Protect tab 3. Connect the load to the DC power under test when the power is OFF. 4. Set a sink current in CC mode of the load. 5. Turn on the power output and the load input. Delay a period of time, the overpower protection of the load will occur. 6. Close all instruments and finish the test.
SIGLENT Troubleshooting Here are some common failures and their solutions. If the problem persists after following the listed steps, please contact SIGLENT. 1. The instrument cannot start up. (1) Check whether the power is correctly connected. (2) Check whether the power switch at the front panel is really on. (3) Remove the power cord and check whether the voltage selector is at the proper scale, whether the specification of the fuse is correct and whether the fuse is intact.
Chapter 4 Troubleshooting SIGLENT Contact SIGLENT America SIGLENT Technologies America, Inc 6557 Cochran Rd Solon, Ohio 44139 Tel: 440-398-5800 Toll Free:877-515-5551 Fax: 440-399-1211 info@siglent.com www.siglentamerica.com Headquarters SIGLENT TECHNOLOGIES CO., LTD. Blog No.4 & No.5, Antongda Industrial Zone, 3rd Liuxian Road, Bao’an District, Shenzhen, 518101, China. Tel:+ 86 755 3661 5186 Fax:+ 86 755 3359 1582 sales@siglent.com www.siglent.