SDL1000X Programming Guide 43
PROGRAM test list. Its unit is s
Example
:SOURce:PROGram:TIME:OFF 4,0.500
Command
Format
[:SOURce]:PROGram:TIME:OFF? <step>
Description
Query the unloading time of the step which is set in this command
in PROGRAM test list. Its unit is s
Example
:SOURce:PROGram:TIME:OFF? 6
Response
2.000
Command
Format
[:SOURce]:PROGram:TIME:DELay <step, {< value > | MINimum
| MAXimum | DEFault}>
Description
Sets test delay time of the step which is set in this command in
PROGRAM test list. Its unit is s
Example
:SOURce:PROGram:TIME: DELay 3,0.500
Command
Format
[:SOURce]:PROGram:TIME:DELay? <step>
Description
Query test delay time of the step which is set in this command in
PROGRAM test list. Its unit is s
Example
:SOURce:PROGram:TIME: DELay? 1
Response
0.600
Command
Format
[:SOURce]:PROGram:MIN <step, {< value > | MINimum |
MAXimum | DEFault}>
Description
Sets the minimum allowed value of the step which is set in this
command in PROGRAM test list. The value is current value in CV
and is voltage value in CC/CR/CP/LED.
Example
:SOURce:PROGram:MIN 3,1.000
Command
Format
[:SOURce]:PROGram:MIN? <step>
Description
Query the minimum allowed value of the step which is set in this
command in PROGRAM test list. The value is current value in CV
and is voltage value in CC/CR/CP/LED.
Example
:SOURce:PROGram:MIN? 5
Response
2.000
Command
Format
[:SOURce]:PROGram:MAX <step, {< value > | MINimum |
MAXimum | DEFault}>