Programming Guide

SIGLENT
SDL1000X Programming Guide 42
Description
Whether set the step of the electronic load which is set in this
command to be short circuit in PROGRAM test list.
Example
:SOURce:PROGram: SHORt 3,ON
Command
Format
[:SOURce]:PROGram:SHORt? <step>
Description
Query Whether the step of the electronic load which is set in this
command is short circuit in PROGRAM test list.
Example
:SOURce:PROGram: SHORt? 4
Response
0
Command
Format
[:SOURce]:PROGram:PAUSE <step, {ON | OFF | 0 | 1 }>
Description
Whether pause the step of the electronic load which is set in this
command in PROGRAM test list.
Example
:SOURce:PROGram: PAUSE 2,1
Command
Format
[:SOURce]:PROGram:PAUSE? <step>
Description
Query Whether the step of the electronic load which is set in this
command is paused in PROGRAM test list.
Example
:SOURce:PROGram: PAUSE? 5
Response
0
Command
Format
[:SOURce]:PROGram:TIME:ON <step, {< value > | MINimum |
MAXimum | DEFault}>
Description
Sets the loading time of the step which is set in this command in
PROGRAM test list. Its unit is s
Example
:SOURce:PROGram:TIME:ON 4,1.000
Command
Format
[:SOURce]:PROGram:TIME:ON? <step>
Description
Query the loading time of the step which is set in this command in
PROGRAM test list. Its unit is s
Example
:SOURce:PROGram:TIME:ON? 5
Response
0.500
Command
Format
[:SOURce]:PROGram:TIME:OFF <step, {< value > | MINimum |
MAXimum | DEFault}>
Description
Sets the unloading time of the step which is set in this command in