Programming Guide

SIGLENT
SDL1000X Programming Guide 41
Format
Description
Sets the current range of the step which is set in this command in
PROGRAM test list.
Example
:SOURce:PROGram:IRANGe 4,5
Command
Format
[:SOURce]:PROGram:IRANGe? <step>
Description
Query the current range of the step which is set in this command in
PROGRAM test list.
Example
:SOURce:PROGram:IRANGe? 5
Response
30
Command
Format
[:SOURce]:PROGram:VRANGe <step, value>
Description
Sets the voltage range of the step which is set in this command in
PROGRAM test list.
Example
:SOURce:PROGram:VRANGe 3,36
Command
Format
[:SOURce]:PROGram:VRANGe? <step>
Description
Query the voltage range of the step which is set in this command in
PROGRAM test list.
Example
:SOURce:PROGram:VRANGe? 2
Response
150
Command
Format
[:SOURce]:PROGram:RRANGe <step, {LOW | MIDDLE | HIGH
| UPPER}>
Description
Sets the resistor range of the step which is set in this command in
PROGRAM test list.
Example
:SOURce:PROGram: RRANGe 3,LOW
Command
Format
[:SOURce]:PROGram:RRANGe? <step>
Description
Query the resistor range of the step which is set in this command in
PROGRAM test list.
Example
:SOURce:PROGram: RRANGe? 3
Response
MIDDLE
Command
Format
[:SOURce]:PROGram:SHORt <step, {ON | OFF | 0 | 1 }>