Programming Guide
SIGLENT
SDL1000X Programming Guide 19
Format
MAXimum | DEFault}
Description
Sets the B Level of CV mode in transient operation.
Example
:SOURce:VOLTage:TRANsient:BLEVel MAXimum
Command
Format
[:SOURce]: VOLTage:TRANsient:BLEVel?
Description
Query the B Level of CV mode in transient operation.
Example
:SOURce: VOLTage:TRANsient:BLEVel?
Response
10.000
Command
Format
[:SOURce]:VOLTage:TRANsient:AWIDth {<value> | MINimum |
MAXimum | DEFault}
Description
Sets the A Level pulse width time value of CV mode in transient
operation. Its unit is “s”
Example
:SOURce:VOLTage:TRANsient:AWIDth DEFault
Command
Format
[:SOURce]:VOLTage:TRANsient:AWIDth?
Description
Query the A Level pulse width time value of CV mode in transient
operation. Its unit is “s”
Example
:SOURce:VOLTage:TRANsient:AWIDth?
Response
1.000
Command
Format
[:SOURce]:VOLTage:TRANsient:BWIDth {<value> | MINimum |
MAXimum | DEFault}
Description
Sets the B Level pulse width time value of CV mode in transient
operation. Its unit is “s”
Example
:SOURce:VOLTage:TRANsient:BWIDth 0.800
Command
Format
[:SOURce]:VOLTage:TRANsient:BWIDth?
Description
Query the B Level pulse width time value of CV mode in transient
operation. Its unit is “s”
Example
:SOURce:VOLTage:TRANsient:BWIDth?
Response
1.500
3.3.4 Source Power Subsystem Command
Command
Format
[:SOURce]:POWer[:LEVel][:IMMediate] {<value> | MINimum |
MAXimum | DEFault}