Programming Guide

SIGLENT
SDL1000X Programming Guide 17
3.3.3 Source Voltage Subsystem Command
Command
Format
[:SOURce]:VOLTage[:LEVel][:IMMediate] {<value> | MINimum |
MAXimum | DEFault}
Description
Sets the preset voltage value of CV mode in static operation.
Example
:SOURce:VOLTage:LEVel:IMMediate 3.000
Command
Format
[:SOURce]:VOLTage[:LEVel][:IMMediate]?
Description
Query the preset voltage value of CV mode in static operation.
Example
:SOURce:VOLTage:LEVel:IMMediate?
Response
4.000
Command
Format
[:SOURce]:VOLTage:IRANGe <value>
Description
Sets the current range of CV mode in static operation.
Example
:SOURce:VOLTage:IRANGe 10
Command
Format
[:SOURce]:VOLTage:IRANGe?
Description
Query the current range of CV mode in static operation.
Example
:SOURce:VOLTage:IRANGe?
Response
30
Command
Format
[:SOURce]:VOLTage:VRANGe <value>
Description
Sets the voltage range of CV mode in static operation.
Example
:SOURce:VOLTage:VRANGe 60
Command
Format
[:SOURce]:VOLTage:VRANGe?
Description
Query the voltage range of CV mode in static operation.
Example
:SOURce:VOLTage:VRANGe?
Response
36
Command
Format
[:SOURce]:VOLTage:TRANsient:MODE {CONTinuous | PULSe |
TOGGle}
Description
Sets the waveform mode of CV mode in transient operation.
Example
:SOURce:VOLTage:TRANsient:MODE TOGGle