Programming Guide

SIGLENT
SDL1000X Programming Guide 16
Description
Query the A Level pulse width time value of CC mode in transient
operation. Its unit is s
Example
:SOURce:CURRent:TRANsient:AWIDth?
Response
1.000
Command
Format
[:SOURce]:CURRent:TRANsient:BWIDth {<value> | MINimum |
MAXimum | DEFault}
Description
Sets the B Level pulse width time value of CC mode in transient
operation. Its unit is s
Example
:SOURce:CURRent:TRANsient:BWIDth 0.800
Command
Format
[:SOURce]:CURRent:TRANsient:BWIDth?
Description
Query the B Level pulse width time value of CC mode in transient
operation. Its unit is s
Example
:SOURce:CURRent:TRANsient:BWIDth?
Response
1.500
Command
Format
[:SOURce]:CURRent:TRANsient:SLEW:POSitive {<value> |
MINimum | MAXimum | DEFault}
Description
Sets the rise slope of CC mode in transient operation.
Example
:SOURce:CURRent:TRANsient: SLEW:POSitive 0.400
Command
Format
[:SOURce]:CURRent:TRANsient:SLEW:POSitive?
Description
Query the rise slope of CC mode in transient operation.
Example
:SOURce:CURRent:TRANsient: SLEW:POSitive?
Response
0.600
Command
Format
[:SOURce]:CURRent:TRANsient:SLEW:NEGative {<value> |
MINimum | MAXimum | DEFault}
Description
Sets the descending slope of CC mode in transient operation.
Example
:SOURce:CURRent:TRANsient: SLEW: NEGative 0.300
Command
Format
[:SOURce]:CURRent:TRANsient:SLEW:NEGative?
Description
Query the descending slope of CC mode in transient operation.
Example
:SOURce:CURRent:TRANsient: SLEW: NEGative?
Response
0.900