Programming Guide
SIGLENT
SDL1000X Programming Guide 11
Command
Format
MEASure:WAVEdata? {CURRent | VOLTage | POWer |
RESistance}
Description
Gets the waveform data of the waveform display interface in
CC/CV/CP/CR mode. Totally include 200 float data .
Example
MEASure:WAVEdata? VOLTage
Response
3.947389,3.928473,3.197493,3.573992…
3.3 Source Subsystem Command
3.3.1 Source Common Subsystem Command
Command
Format
[:SOURce]:INPut[:STATe] {ON | OFF | 0 | 1}
Description
Sets the input status of the load (ON or OFF)
Example
:SOURce:INPut:STATe OFF
Command
Format
[:SOURce]:INPut[:STATe]?
Description
Query the input status of the load. Return “1” if input status is ON.
Otherwise, return “0”
Example
:SOURce:INPut:STATe?
Response
0
Command
Format
[:SOURce]:SHORt[:STATe] {ON | OFF | 0 | 1}
Description
Sets the short circuit status of the load (ON or OFF)
Example
:SOURce:SHORt:STATe ON
Command
Format
[:SOURce]:SHORt[:STATe]?
Description
Query the short circuit status in current mode of the load. Return
“1” if short circuit status is ON. Otherwise, return “0”
Example
:SOURce:SHORt:STATe?
Response
1
Command
Format
[:SOURce]:FUNCtion:TRANsient {CURRent | VOLTage | POWer
| RESistance}
Description
Sets mode in transient operation (CC/CV/CP/CR)