Programming Guide
SIGLENT
SDL1000X Programming Guide 9
Response
0
Command
Format
*OPC
Description
Set bit 0 in the standard event status register to “1” when all
operations have finished.
Example
*OPC
Command
Format
*OPC?
Description
Query whether the current operation has been finished. Returns a
“1” if the operation is complete.
Example
*OPC?
Response
1
Command
Format
*SRE <number>
Description
Set the bits in the status byte enable register.
Example
*SRE 24
Command
Format
*SRE?
Description
Query the status byte enable register. The value returned reflects
the current state of all the bits in the register.
Example
*SRE?
Response
24
Command
Format
*STB?
Description
Query the status byte event register. The value returned reflects the
current state of all the bits in the register.
Example
*STB?
Response
72
Command
Format
*TST?
Description
This query is used by some instruments for a self test.
Example
*TST?
Response
0