Programming Guide
SIGLENT
SDL1000X Programming Guide 7
written completely. For example:
:SOURce:CURRent:SLEW:POSitive?
Can be abbreviated to:
:SOUR:CURR:SLEW:POS?
2.5 Special instructions
1、 When sets the current range,if <value> is greater than 5, the current range
will be set to 30A; if <value> is less than 5, the current range will be set to
5A.
2、 When sets the voltage range,if <value> is greater than 36, the voltage
range will be set to 150V; if <value> is less than 36, the voltage range will
be set to 36V.
3、 The parameters is {<value> | MINimum | MAXimum | DEFault}
enumerations. Its meanings are as follows
<value> :Any setting value
MINimum :The minimum value in the setting range
MAXimum :The maximum value in the setting range
DEFault :The default value for the setting value