Data Sheet

SDL1000X Programmable DC Electronic Load
WWW.SIGLENT.COM
Model SDL1020X SDL1020X-E
Input Rating
(0~40
)
Input voltage 0~150 V
Input current 0~5 A 0~30 A 0~5 A 0~30 A
Input power 200 W
Minimum Operating
Voltage (DC)
0.15 V at 5 A 0.9 V at 30 A 0.15 V at 5 A 0.9 V at 30 A
CV Mode
Range 0~36 V 0~150 V 0~36 V 0~150 V
Resolution 1 mV 1 mV
Accuracy ± (0.05%+0.025%FS) 50 ppm/
± (0.05%+0.025%FS) 50 ppm/
CC Mode
Range 0~5 A 0~30 A 0~5 A 0~30 A
Resolution 1 mA 1 mA
Accuracy *2 ± (0.05%+0.05%FS) 100 ppm/
± (0.1%+0.1%FS) 100 ppm/
CR Mode *1
Range 0.03 Ω~10 kΩ
Resolution 1mA/Vsense
Accuracy 0.01%+0.0008 s [1]
CP Mode *3
Range 200 W
Resolution 10 mW
Accuracy 0.1%+0.1% FS
Dynamic Model
CC Mode
T1&T2 20 us~999 S/Res: 1 uS 20 us~999 S/Res: 1 uS
Accuracy 20 us~200 ms/Acc: 5 us, 200 ms ~999 s/Acc: 5 ms 20 us~200 ms/Acc: 5 us, 200 ms ~999 s/Acc: 5 ms
Current Slew Rate *4 0.001~0.5 A/us 0.001~2.5 A/us 0.001~0.5 A/us 0.001~2.5 A/us
Minimum Rise Time ≈10 uS ≈12 uS ≈10 uS ≈12 uS
Measuring Range
Readback Voltage
Range 0~36 V 0~150 V 0~36 V 0~150 V
Resolution 0.1 mV 1 mV 1 mV
Accuracy ± (0.025%+0.025% FS) 20 ppm/
± (0.05%+0.02% FS) 20 ppm/
Readback Current
Range 0~5 A 0~30 A 0~5 A 0~30 A
Resolution 0.1 mA 1 mA
Accuracy ± (0.05%+0.05% FS) 50 ppm/
± (0.05%+0.05% FS) 50 ppm/
Readback Power
Range 200 W
Resolution 10 mW
Accuracy ± (0.1%+0.1% FS)
Readback Resistance
Range 0.03 Ω ~10 KΩ
Resolution 0.001Ω
Specications
Unless otherwise noted, all specications are guaranteed within the temperature range of 25°C ± 5 °C with warm-up time of 30 minutes.