1. To Set the Output Load
For the [Output] connector on the front panel, the generator has a built-in
50Ω series impendence. If the actual load does not match the set one,
the displayed amplitude and offset will be incorrect. This function is used
to match the displayed voltage with the expected one.
Steps for setting the load:
Press Utility →Output Setup→ Load, to enter the following interface.
Please note that the load parameter shown on the right
bottom is the default setting when the power is on or the pre-set load
value. If the current value is valid for the output, then current value will be
used.
Figure 2- 59 Set the Output Load
Instruction
SDG1000 Series has a fixed 50Ω Series Impendence. No matter what value
the set parameter is, if the real load is different from the set one, the displayed
voltage will not equal the real voltage.