3. Apply the gas listed on the Data Label to the inlet at the recommended pressure (listed on the Data
Label/Calibration Certificate). Your 50 Series will begin to measure mass flow. This will be
displayed on the user software or local display.
4. You will need a setpoint input in order to control flow. Do so carefully! You can digitally input a
setpoint using the user software or control directly with your analog setpoint source if you ordered
the analog option. NOTE: The unit ships in the configuration ordered. This information is also listed
on the Data Label and on the Calibration Certificate. This information is also inherent in the model
code. See the Instruction Manual for more details.
5. DO NOT leave a setpoint applied for an extended period of time to a controller when the gas supply
is off or blocked. Damage may result from excessive heating, and the unit will become hot
enough to burn you. In this case you should adjust your setpoint to 0.000 or power down the unit.
WIRING THE OPTIONAL COMMS CABLE
(Pin-out)
USER SOFTWARE INSTALLATION
Each Smart-Trak
®
50 Series order is shipped with a CD-ROM containing the Smart-Trak
®
50 Series
Software. This software allows you to view the mass flow rate, full scale flow rate value, and setpoint.
Use the software to zero the instrument, revert to the default zero, change the span, change setpoint or
power-up setpoint, or change the address of the unit software is talking to (RS-485 only). Locate this
disk. At this point, EXIT OUT OF ANY OPEN APPLICATIONS BEING RUN ON YOUR
COMPUTER.
1. Insert the Smart-Trak
®
50 Series Software CD into your CD-ROM
2. Open “My Computer” on your desktop
3. Open the CD Named: “Smart-Trak
®
50 Series” on your D-drive
4. Run “setup.exe”
5. Follow the instructions on screen
RUNNING THE USER SOFTWARE
Locate the program named “50 Series User Software” and open it. You will see the following main
screen. Select ComPort in the menu and follow the instructions below for setup:
Pin #
Color Function
1. White
4-20 mA / 0-5 VDC Output
2. Brown
4-20 mA / 0-5 VDC Output Return
3. Blue
4-20 mA / 0-5 VDC Setpoint in
4. Orange
4-20 mA / 0-5 VDC Setpoint Return
5. Red
+24 VDC in/Power Input (+)
6. Green
Power Return (-)/RS 232 Ground
7. Yellow
N/C (RS-485 Bus Isolated Ground)
8. Purple
RS-232 Tx (RS-485 Bus B, +, D1)
9. Gray
RS-232 Rx (RS-485 Bus A, -, D0)
Shield Shield Earth Ground
Figure: DB9 Pin-out (There are two
DB9 connectors, one on each side of the
enclosure. Assure block off plate covers
unused DB9.