Instruction manual
Instruction Manual Smart-Trak
®
50 Series
IM-50 Rev. A.2 36
Appendix B: Product Specifications
Performance Specifications
Accuracy
Air, N2, O2, CO2, CO, Ar: +/-1.5 % of full scale including linearity over 32
°
F to 122
°
F (0
°
C to
50
°
C) and 5 to 145 psig (0.3 to 10 barg)
Other gases: +/- 3 of full scale including linearity at operating conditions (consult factory)
Repeatability
+/- 0.25% of full scale
Temperature Coefficient
0.025% of full scale per
°
F (0. 05% of full scale per
°
C), or better
Pressure Coefficient
0.01% of full scale per psi (0.15% of full scale per bar), or better
Response Time
300 ms time constant; two seconds (typical) to within +/- 2% of final value including setting time
May be tuned to be faster or slower (consult factory)
Operating Specifications
Gases
All neutral, non-contaminated dry and clean gases (air, nitrogen, carbon dioxide, argon, methane,
hydrogen, helium, etc.) compatible with wetted materials aluminum, stainless steel and Viton
®
Mass Flow Rates
From 0 to 20 sccm or nml/m to 0 to 50 slpm or nlpm; flow ranges specified are for an equivalent
flow of air at 760 mm Hg and 21
°
C (70
°
F) for standard conditions, 760 mm Hg. and 0
°
C (32
°
F) for
normal conditions, other gases will have other ranges (consult factory); other units are available
(e.g., scfh or nm3/h).
F1 or N1: 20 sccm or nccm. Minimum flow is zero. You must specify full scale. Must fall within
listed range.
F2 or N2: 200 sccm or nccm, F3 or N3: 1.0 slpm or nlpm, F4 or N4: 5.0 slpm or nlpm, F5 or
N5: 10.0 slpm or nlpm, F6 or N6: 20.0 slpm of nlpm, F7 or N7: 50.0 slpm or nlpm.
Other ranges in other units are available (e.g., scfh or nm3/h), but may extend lead time (consult
factory).