4. Close the window using the X button in the upper right corner of the dialog box. The
communications are set up, and the unit should be communicating with the computer, as evident
by the changing numbers right of the COM1: port in the picture below. If the numbers are not
changing, click ‘Get Parameters.’
RESET NOTE: If you are not able to establish communication with your instrument be sure you have the proper
ComPort selected to match your PC port then hit the “Reset Host Port To Default” button in the Port Dialog Box and
power down your instrument. Next, power up the instrument again and hold the zero pushbutton (located inside your
instruments side plate behind the silver socket head cap screw) for a minimum of one second after powering up your
instrument. This will reset both the host software and your instrument to the default settings. Click on the ‘Get
Parameters’ button and communication should be established.